  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Chester",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62749",
      "recalling_firm": "Merit Medical Systems, Inc.",
      "address_1": "12701 N Kingston Ave",
      "address_2": "N/A",
      "postal_code": "23836-2700",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution: US including states of: NE and SD.",
      "recall_number": "Z-0464-2013",
      "product_description": "Custom Procedural Trays: Merit's Custom Total Hip Pack, K12T-05537, Merit's Total Knee Pack, K12T05538    A custom procedural tray is a combination of legally marketed medical devices placed into one container for the convenience of the user and used within the limits of the product's intended uses.",
      "product_quantity": "83 packs",
      "reason_for_recall": "Custom Procedural trays contain Stryker Hytrel Togas which are being recalled because the clear tape that is intended to aid in the prevention of patient's fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process.",
      "recall_initiation_date": "20120608",
      "center_classification_date": "20121203",
      "termination_date": "20121210",
      "report_date": "20121212",
      "code_info": "Merit's Custom Total Hip Pack, K12T-05537, Lot  numbers T320054, T331364, T335090; Merit's Total Knee Pack, K12T05538, Lot numbers T320059, T331365, T335092"
    }
  ]
}