  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Des Plaines",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69542",
      "recalling_firm": "Abbott Molecular",
      "address_1": "1300 E Touhy Ave",
      "address_2": "N/A",
      "postal_code": "60018-3315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US nationwide (CA, CT, FL, NC, PA, SC, and DC), Algeria\t,  Australia\t,  Austria\t,  Belgium\t,  Bosnia and Herzegovina\t,  Cameroon\t,  Canada\t,  Chad\t,  Colombia\t,  Costa Rica\t,  Croatia\t,  Denmark\t,  Egypt\t,  Ethiopia\t,  Finland\t,  France\t,  Germany\t,  Greece\t,  Guinea\t,  Honduras\t,  Hong Kong\t,  India\t,  Indonesia\t,  Ireland\t,  Israel\t,  Italy\t,  Japan\t,  Kenya\t,  Korea\t,  Lithuania\t,  Mali\t,  Mexico  \t,  Netherlands\t,  New Zealand\t,  Nigeria\t,  Norway\t,  Pakistan\t,  Paraguay\t,  Poland\t,  Portugal\t,  Romania\t,  Russia\t,  Saudi Arabia\t,  Senegal\t,  Slovenia\t,  South Africa\t,  Spain\t,  Sweden\t,  Switzerland\t,  Taiwan\t,  Tanzania\t,  Thailand\t,  Turkey\t,  Uganda\t,  Ukraine\t,  United Arab Emirates,   United Kingdom\t,  Vietnam",
      "recall_number": "Z-0463-2015",
      "product_description": "Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.",
      "product_quantity": "9 CD ROMs (US); 1,273 CD ROMs (Foreign)",
      "reason_for_recall": "Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.",
      "recall_initiation_date": "20141023",
      "center_classification_date": "20141125",
      "termination_date": "20160308",
      "report_date": "20141203",
      "code_info": "List Number: 09K14; Models: E series, G series",
      "more_code_info": ""
    }
  ]
}