{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Des Plaines",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89168",
      "recalling_firm": "Abbott Molecular, Inc.",
      "address_1": "1300 E Touhy Ave",
      "address_2": "N/A",
      "postal_code": "60018-3315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US distribution and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozambique, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Uganda, United Kingdom, and Vietnam",
      "recall_number": "Z-0462-2022",
      "product_description": "Alinity m System, Part No. 08N53-002",
      "product_quantity": "742 devices",
      "reason_for_recall": "There is a software defect that can cause the Amplification Detection clamp CLEAN position to be set too high.",
      "recall_initiation_date": "20211206",
      "center_classification_date": "20220106",
      "report_date": "20220112",
      "code_info": "Alinity m System, Model 08N53-002, all serial numbers"
    }
  ]
}