  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brookfield",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72738",
      "recalling_firm": "Siemens Healthcare Diagnostics Inc.",
      "address_1": "101 Silvermine Rd",
      "address_2": "N/A",
      "postal_code": "06804-2047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of   Canada and Mexico.",
      "recall_number": "Z-0460-2016",
      "product_description": "Reagent Management System (RMS)  Dimension RxL Max w/o HM   Product Usage:  The Reagent Management System (RMS) is a module for use with the Dimension¿ clinical chemistry system. The RMS doubles the available storage of refrigerated Flex¿ reagent cartridges on the Dimension¿ system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex¿ reagent cartridges, allowing the Dimension¿ system to process samples without interruption.",
      "product_quantity": "3",
      "reason_for_recall": "Siemens has  identified RMS refrigeration compressors that are missing the safety cover on the electrical  termination block. The exposed electrical hazard on the electrical termination block is located  directly behind the RMS Waste Container.",
      "recall_initiation_date": "20151110",
      "center_classification_date": "20151221",
      "termination_date": "20170823",
      "report_date": "20151230",
      "code_info": "Dimension RxL Max w/o HM Catalog Numbers 752000.901 (Domestic)/752000.921 (Intl).  RxL Max HM Catalog Numbers 752000.911 (Domestic)/752000.931 (Intl).  All serial numbers."
    }
  ]
}