  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Schwarzenbruck",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66808",
      "recalling_firm": "Iba Dosimetry Gmbh",
      "address_1": "Bahnhofstr. 5",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany)  Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.",
      "recall_number": "Z-0456-2014",
      "product_description": "iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator)    radiological",
      "product_quantity": "33 units",
      "reason_for_recall": "The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.",
      "recall_initiation_date": "20130820",
      "center_classification_date": "20131206",
      "termination_date": "20140224",
      "report_date": "20131218",
      "code_info": "SW Version 3.0. Catalong Number CS10-100",
      "more_code_info": ""
    }
  ]
}