  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72632",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "address_2": "N/A",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution.",
      "recall_number": "Z-0455-2016",
      "product_description": "Alaris EtCO2 Module, Model 8300.  A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).",
      "product_quantity": "157 units",
      "reason_for_recall": "CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2.  The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.",
      "recall_initiation_date": "20151112",
      "center_classification_date": "20151221",
      "termination_date": "20170123",
      "report_date": "20151230",
      "code_info": "9905776  9905792  12331254  12350410  12352410  12364342  12364343  12364901  12369088  12401328  12402583  12479923  12481239  12593721  12623833  12623957  12668253  12679872  12697242  12698542  12736367  12805782  12807757  12868325  12870021  12916518  12918611  12919079  12921157  13034011  13034375  13038124  13050300  13050648  13051034  13061112  13072829  13075412  13077962  13088823  13106029  13106030  13106138  13106140  13122164  13122401  13122402  13122936  13136786  13141565  13142603  13143660  13143809  13143889  13167255  13169672  13173261  13177210  13178248  13178389  13178391  13178512  13178968  13178969  13186760  13187737  13188106  13189792  13190670  13216601  13220274  13316434  13316444  13316453  13316473  13316496  13317481  13317552  13317553  13457448  13463929  13464899  13469917  13470004  13470021  13476713  13476912  13477643  13480785  13482441  13482518  13489475  13491251  13521218  13559325  13559686  13569291  13582129  13602388  13621652  13621847  13622337  13624519  13628251  13628876  13628877  13628918  13629153  13658756  13732201  13744088  13781347  13782359  13796475  13809762  13813539  13818275  13853185  13853375  13858182  13858798  13858976  13862567  13867168  13895850  13895989  13896107  13896108  13896110  13896347  13896456  13896669  13896747  13896748  13910604  13922145  13951711  13994625  13995332  14000190  14006401  14035430  14036040  14084571  14084629  14126982  14126990  14151686  14152555  14172846  14190913  14228679  14235981  14253982  14258801  14259542  14280651"
    }
  ]
}