{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63634",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and Pacific countries, and the countries of  EEMEA, Canada, Latin America, and Japan.",
      "recall_number": "Z-0455-2013",
      "product_description": "Neptune¿ 2 Rover - ULTRA ~230V ~ 50 Hz 12A  REF 0702-002-000, Rx Caution, consult accompanying documents  Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg,     The Neptune Waste Management System consists of a mobile rover unit  used to suction and collect fluid waste, small debris and electrocautery  smoke from a surgical site. Mobility allows the rover unit to be  relocated to a waste disposal area where the rover\u0019s collection canisters  can be emptied, via the docking station.    Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices.",
      "product_quantity": "154",
      "reason_for_recall": "Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.",
      "recall_initiation_date": "20110524",
      "center_classification_date": "20121203",
      "termination_date": "20140528",
      "report_date": "20121212",
      "code_info": "Part 0702-002-000 (230v)     Lot 0733700353 through 1036100423"
    }
  ]
}