  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raynham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72654",
      "recalling_firm": "DePuy Spine, Inc.",
      "address_1": "325 Paramount Dr",
      "address_2": "N/A",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : NC, MI, CO, MD, FL, ID, TX, CA and NY.",
      "recall_number": "Z-0454-2016",
      "product_description": "Aegis Torque Handle CATALOG NO. : 2871-10-800.  A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.",
      "product_quantity": "16 devices",
      "reason_for_recall": "Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length",
      "recall_initiation_date": "20151120",
      "center_classification_date": "20151221",
      "termination_date": "20170703",
      "report_date": "20151230",
      "code_info": "LOT NUMBER: GB0906 GB0707 GB0907 GB58815 GB59244 GB61134"
    }
  ]
}