      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69286",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and  RUSSIA.",
      "recall_number": "Z-0451-2015",
      "product_description": "Innova IGS 540    Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.",
      "product_quantity": "10 (4 US, 6 OUS).",
      "reason_for_recall": "Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.",
      "recall_initiation_date": "20140904",
      "center_classification_date": "20141212",
      "termination_date": "20150121",
      "report_date": "20141224",
      "code_info": "Mfg Lot or Serial # System ID  00000642076BU4 303695IR1  00000642193BU7 0004040462  00000646910BU0 314768SM10  00000647792BU1 904UFHNSP  00000646492BU9 XV386041GE8  00000646048BU9 S4160099  00000644611BU6 H2009VAS44  00000646493BU7 H0008VAS06  00000645027BU4 XV106184CX5  00000646491BU1 4184240"
    }
  ]
}