  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72741",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR  CHILE, COSTA RICA, NETHERLANDS, CANADA",
      "recall_number": "Z-0450-2016",
      "product_description": "A.L.P.S. 4.0mm Cancellous Locking Screws   intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.",
      "product_quantity": "182, 965",
      "reason_for_recall": "These screws have a single lead (instead of a triple lead) thread.  The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.",
      "recall_initiation_date": "20151120",
      "center_classification_date": "20151218",
      "termination_date": "20160908",
      "report_date": "20151230",
      "code_info": "Part number and size (non-sterile)  816140010, 4.0X10MM;  816140012, 4.0X12MM;  816140014, 4.0X14MM;  816140016, 4.0X16MM;  816140018, 4.0X18MM;  816140020, 4.0X20MM;  816140022, 4.0X22MM;  816140024, 4.0X24MM;  816140026, 4.0X26MM;  816140028, 4.0X28MM;  816140030, 4.0X30MM;  816140032, 4.0X32MM;  816140034, 4.0X34MM;  816140036, 4.0X36MM;  816140038, 4.0X38MM;  816140040, 4.0X40MM;  816140042, 4.0X42MM;  816140044, 4.0X44MM;  816140046, 4.0X46MM;  816140048, 4.0X48MM;  816140050, 4.0X50MM;  816140055, 4.0X55MM;  816140060, 4.0X60MM;  816140065, 4.0X65MM;  816140070, 4.0X70MM;  816140075, 4.0X75MM;  816140080, 4.0X80MM;  816140085, 4.0X85MM;  816140090, 4.0X90MM;  816140095, 4.0X95MM;  816140100, 4.0X100MM    Part number and size (sterile)  856140010, 4.0X10MM;  856140012, 4.0X12MM;  856140014, 4.0X14MM;  856140016, 4.0X16MM;  856140018, 4.0X18MM;  856140020, 4.0X20MM;  856140022, 4.0X22MM;  856140024, 4.0X24MM;  856140026, 4.0X26MM;  856140028, 4.0X28MM;  856140030, 4.0X30MM;  856140032, 4.0X32MM;  856140034, 4.0X34MM;  856140036, 4.0X36MM;  856140038, 4.0X38MM;  856140040, 4.0X40MM;  856140042, 4.0X42MM;  856140044, 4.0X44MM;  856140046, 4.0X46MM;  856140048, 4.0X48MM;  856140050, 4.0X50MM;  856140055, 4.0X55MM;  856140060, 4.0X60MM;  856140065, 4.0X65MM;  856140070, 4.0X70MM.  All lots manufactured between 06/01/08 and 06/01/15."
    }
  ]
}