{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69638",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.",
      "recall_number": "Z-0447-2015",
      "product_description": "8MM, CURVED BIPOLAR DISSECTOR;    Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.",
      "product_quantity": "509 total units, all affected devices",
      "reason_for_recall": "Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.",
      "recall_initiation_date": "20141103",
      "center_classification_date": "20141125",
      "termination_date": "20150203",
      "report_date": "20141203",
      "code_info": "Part number; 470344-04;  All 8mm da Vinci Xi (IS4000) Surgical System instruments  manufactured with white colored housings.",
      "more_code_info": ""
    }
  ]
}