{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Shanghai",
      "state": "",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86269",
      "recalling_firm": "Shanghai United Imaging Healthcare Co., Ltd.",
      "address_1": "No. 2258 Chengbei Rd, Jiading Ind.",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution including in the state of TX.",
      "recall_number": "Z-0446-2021",
      "product_description": "Computed Tomography X-Ray System, Model uCT 550, Rx, CE, GTIN:  06971576831036 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).",
      "product_quantity": "5",
      "reason_for_recall": "A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during  multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.",
      "recall_initiation_date": "20200805",
      "center_classification_date": "20201112",
      "report_date": "20201118",
      "code_info": "Serial Numbers:  353020, 353021, 353023, 353045, 353049",
      "more_code_info": ""
    }
  ]
}