{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Elyria",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89062",
      "recalling_firm": "Invacare Corporation",
      "address_1": "1200 Taylor St",
      "address_2": "N/A",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.",
      "recall_number": "Z-0445-2022",
      "product_description": "Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A)  Models: TDX SP2: TDXSP2, TDXSP2-CG, TDXSP2X-CG, TDXSP2-MCG, TDXSP2X-MCG, TDXSP2V, TDXSP2V-HD, TDXSP2HD, TDXSP2HD-CG, TDXSP2HD-MCG",
      "product_quantity": "1477 units US; 1226 units OUS",
      "reason_for_recall": "Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury",
      "recall_initiation_date": "20211112",
      "center_classification_date": "20220103",
      "report_date": "20220112",
      "code_info": "LiNX Gyro firmware version 6.1.2  UDI-DI Numbers:       TDXSP2  00841447100911     TDXSP2-CG  00841447100942     TDXSP2-MCG  00841447100966     TDXSP2HD  00841447108641     TDXSP2HD-CG  00841447111207     TDXSP2HD-MCG  00841447111214     TDXSP2V  00841447101512     TDXSP2V-HD  00841447102052     TDXSP2X-CG  00841447101000     TDXSP2X-MCG  00841447101017"
    }
  ]
}