      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89228",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of IA, LA, NE, NJ, NV, NY, OH, PA, SC, SD.",
      "recall_number": "Z-0441-2022",
      "product_description": "ARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.  Model Number: 8162815",
      "product_quantity": "13 units",
      "reason_for_recall": "Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury",
      "recall_initiation_date": "20211209",
      "center_classification_date": "20211229",
      "termination_date": "20230504",