  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Northridge",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72352",
      "recalling_firm": "Medtronic MiniMed Inc.",
      "address_1": "18000 Devonshire St",
      "address_2": "N/A",
      "postal_code": "91325-1219",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Internationally Distribution only to the countries of :  Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.",
      "recall_number": "Z-0437-2016",
      "product_description": "MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710    The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.",
      "product_quantity": "53 units",
      "reason_for_recall": "The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.",
      "recall_initiation_date": "20150925",
      "center_classification_date": "20151217",
      "termination_date": "20160205",
      "report_date": "20151223",
      "code_info": "NG1019903H  NG1019906H  NG1019907H  NG1019909H  NG1019914H  NG1019915H  NG1019916H  NG1019918H  NG1019921H  NG1019922H  NG1019923H  NG1019925H  NG1019926H  NG1019928H  NG1019930H  NG1019931H  NG1019932H  NG1019933H  NG1019935H  NG1019937H  NG1019939H  NG1019940H  NG1019943H  NG1019944H  NG1019946H  NG1019949H  NG1019950H  NG1019953H  NG1019966H  NG1019971H  NG1019975H  NG1019976H  NG1019979H  NG1019981H  NG1019989H  NG1019997H  NG1019999H  NG1020001H  NG1020006H  NG1020007H  NG1020012H  NG1020015H  NG1020017H  NG1020018H  NG1020019H  NG1020020H  NG1020022H  NG1020023H  NG1020026H  NG1020027H  NG1020028H  NG1020029H  NG1020031H"
    }
  ]
}