  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66866",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and to countries of:  Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Denmark, France, Germany, India, Israel, Italy, Japan, S. Korea, Luxembourg, Mexico, Monaco, Netherlands, Norway, Romania. Russian Federation, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.",
      "recall_number": "Z-0436-2014",
      "product_description": "Assembly, Patient Side Manipulator on da Vinci Surgical Systems:  da Vinci S Surgical System;  da Vinci SI Surgical System  da Vinci Si-e Surgical System;    Manufactured by Intuitive Surgical  Sunnyvale, CA 94086.    The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS2000/3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electro cautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electro cautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system is indicated for adult and pediatric use",
      "product_quantity": "1,386",
      "reason_for_recall": "Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by the surgeon as resistance in the movement of the master.In this situation, the instrument can stall momentarily and then suddenly catch-up to the master position if the surgeon pushes through the resistance.",
      "recall_initiation_date": "20131111",
      "center_classification_date": "20131203",
      "termination_date": "20140218",
      "report_date": "20131211",
      "code_info": "Part number 380900-02 and 380900-03:  System ID numbers - US:   SH0018  SH0206  SH0407  SH0752  SH0981  SH1179  SH1417  SH1690  SH1785  SH1887  SH0022  SH0214  SH0408  SH0755  SH0983  SH1180  SH1418  SH1699  SH1786  SH1889  SH0025  SH0216  SH0409  SH0756  SH0987  SH1186  SH1430  SH1706  SH1787  SH1890  SH0030  SH0223  SH0414  SH0762  SH0989  SH1191  SH1437  SH1707  SH1789  SH1891  SH0032  SH0226  SH0423  SH0765  SH0994  SH1194  SH1441  SH1709  SH1790  SH1892  SH0036  SH0228  SH0427  SH0769  SH0997  SH1199  SH1445  SH1710  SH1791  SH1893  SH0038  SH0232  SH0428  SH0770  SH1007  SH1201  SH1448  SH1711  SH1792  SH1894  SH0040  SH0240  SH0431  SH0772  SH1011  SH1203  SH1459  SH1719  SH1793  SH1896  SH0042  SH0243  SH0436  SH0781  SH1012  SH1215  SH1461  SH1721  SH1794  SH1897  SH0049  SH0245  SH0437  SH0791  SH1013  SH1246  SH1507  SH1727  SH1795  SH1899  SH0053  SH0248  SH0438  SH0808  SH1014  SH1259  SH1517  SH1728  SH1797  SH1900  SH0055  SH0254  SH0439  SH0819  SH1016  SH1269  SH1518  SH1729  SH1798  SH1903  SH0057  SH0258  SH0448  SH0821  SH1024  SH1269  SH1531  SH1731  SH1808  SH1904  SH0060  SH0262  SH0453  SH0823  SH1027  SH1273  SH1539  SH1733  SH1809  SH1905  SH0065  SH0265  SH0454  SH0829  SH1029  SH1283  SH1564  SH1735  SH1810  SH1906  SH0066  SH0266  SH0466  SH0838  SH1030  SH1285  SH1580  SH1736  SH1811  SH1907  SH0067  SH0268  SH0473  SH0845  SH1031  SH1286  SH1581  SH1739  SH1812  SH1908  SH0068  SH0272  SH0475  SH0847  SH1032  SH1288  SH1583  SH1741  SH1814  SH1909  SH0070  SH0275  SH0489  SH0849  SH1034  SH1290  SH1589  SH1743  SH1815  SH1910  SH0076  SH0282  SH0490  SH0850  SH1036  SH1293  SH1594  SH1744  SH1817  SH1912  SH0078  SH0284  SH0493  SH0853  SH1050  SH1294  SH1598  SH1746  SH1822  SH1913  SH0082  SH0292  SH0521  SH0855  SH1053  SH1306  SH1599  SH1747  SH1823  SH1915  SH0085  SH0296  SH0524  SH0858  SH1055  SH1322  SH1605  SH1748  SH1824  SH1916  SH0108  SH0304  SH0525  SH0862  SH1059  SH1324  SH1606  SH1749  SH1827  SH1922  SH0112  SH0309  SH0527  SH0865  SH1069  SH1331  SH1612  SH1750  SH1831  SH1923  SH0113  SH0310  SH0543  SH0866  SH1080  SH1332  SH1617  SH1752  SH1840  SH1924  SH0115  SH0314  SH0548  SH0867  SH1081  SH1333  SH1629  SH1753  SH1841  SH1925  SH0134  SH0318  SH0557  SH0871  SH1082  SH1334  SH1630  SH1754  SH1842  SH1926  SH0141  SH0319  SH0577  SH0873  SH1085  SH1335  SH1631  SH1760  SH1846  SH1933  SH0145  SH0326  SH0585  SH0875  SH1091  SH1336  SH1632  SH1761  SH1847  SH1934  SH0147  SH0327  SH0592  SH0877  SH1096  SH1337  SH1633  SH1763  SH1850  SH1935  SH0149  SH0330  SH0619  SH0883  SH1099  SH1343  SH1634  SH1764  SH1855  SH1942  SH0157  SH0331  SH0633  SH0886  SH1110  SH1347  SH1635  SH1765  SH1856  SH1952  SH0160  SH0332  SH0644  SH0904  SH1120  SH1363  SH1636  SH1766  SH1857  SH1955  SH0165  SH0335  SH0653  SH0905  SH1124  SH1365  SH1645  SH1767  SH1862  SH0167  SH0341  SH0670  SH0920  SH1125  SH1367  SH1646  SH1769  SH1864  SH0171  SH0343  SH0684  SH0925  SH1131  SH1370  SH1650  SH1770  SH1865  SH0172  SH0344  SH0695  SH0928  SH1132  SH1371  SH1651  SH1771  SH1867  SH0180  SH0355  SH0700  SH0932  SH1133  SH1372  SH1655  SH1772  SH1872  SH0181  SH0374  SH0710  SH0936  SH1134  SH1375  SH1658  SH1773  SH1873  SH0182  SH0378  SH0717  SH0942  SH1137  SH1376  SH1665  SH1774  SH1876  SH0183  SH0384  SH0722  SH0946  SH1139  SH1381  SH1669  SH1775  SH1877  SH0184  SH0385  SH0723  SH0949  SH1142  SH1386  SH1670  SH1777  SH1878  SH0187  SH0390  SH0731  SH0962  SH1157  SH1390  SH1673  SH1778  SH1882  SH0190  SH0391  SH0734  SH0966  SH1165  SH1396  SH1679  SH1780  SH1883  SH0196  SH0398  SH0739  SH0972  SH1170  SH1400  SH1680  SH1782  SH1884  SH0198  SH0401  SH0748  SH0973  SH1177  SH1401  SH1683  SH1783  SH1885  SH0202  SH0403  SH0751  SH0976  SH1178  SH1402  SH1689  SH1784  SH1886  USG008  USG077  USG113  USG166  USG013  USG092  USG114  USG168  USG020  USG093  USG136  USG169  USG044  USG095  USG146  USG170  USG050  USG104  USG151  USG172  USG074  USG105  USG156  USG180  SG014  SG163  SG235  SG382  SG457  SG546  SG627  SG699  SG031  SG182  SG247  SG387  SG503  SG555  SG629  SG706  SG049  SG194  SG273  SG408  SG504  SG590  SG633  SG712  SG084  SG2020  SG278  SG437  SG507  SG594  SG641  SG751  SG117  SG2024  SG282  SG442  SG523  SG595  SG688  SG781  SG138  SG205  SG329  SG450  SG528  SG607  SG691  SG801  SG161  SG224  SG358  SG454  SG530  SG613  SG698  SG824  SH0584 SH1603 SH1788  SH0628 SH1656 SH1898 SH0999 SH1657 SH1902 SH1075 SH1658 SH1954 SH1108 SH1667 SH1994 SH1340 SH1692 SH1374 SH1720 SH1578 SH1759.  International:  Australia:  SH0689      SH0690      SH1275      SH1278      SH1486      SH1861      SH1875      SH1879      USG319      USG332;  Austria:  SG579;  Belgium:  SG430      SH0034      SH1405      SH1406      SH1638;  Bulgaria:  SH1692;  Brazil:  SH1768      SG664      SH1866;  Canada:  SG242      SG345      SH0260      SH0370      SH1095;  China:  SG815      SG831      SH0965      SH1062      SH1582;  Columbia:  SH1161;  Denmark:  SH0975      SH1168      SH1455      SH1716      SH1756;  France:  SG140      SH0086      SH0372      SH0546      SH0939      SH0971      SH1075      SH1108      SH1639      SH1686      SH1688      SH1740      SH1843      SH1844      USG323      USG798;  Germany:  SG756  SH0028  SH0593  SH0745  SH0809  SH0924  SH1642  USG106  USG122  USG145  India:  SG817      SH0812      SH0968      SH0969      SH1228      SH1229      SH1762:  Israel:  SH0816;     Italy:  SG489      SH0148      SH0209      SH0329      SH0757      SH0878      SH0991      SH1587      SH1588      SH1880      SH1881      USG103      USG364        USG398        SG258;  Luxembourg:  SH1920;  Mexico:  SH1931        SH1316;  Monaco:  SH0208;  Netherlands:  SH0827        SH0923        SH1579        SH1863;  Norway:  SH0205        SH1159        SH1297        SH1361;  Romania:  SH0161;  Russia:  SH1796        SH1868        SH1869        SH1870        SH1871        SH1936;  Japan:  SG845      SG865      SG893      SG895      SG898      SG901      SH1693      SH1694      SH1696      SH1697      SH1703      SH1704      SH1708      SH1714      SH1722      SH1726      SH1801      SH1802      SH1803      SH1804      SH1819      SH1820      SH1821     SH1829     SH1830     SG862     SH1695     SH1698     SH1700     SH1701     SH1702     SH1705     SH1712     SH1713     SH1715     SH1718     SH1723     SH1724     SH1725     SH1742     SH1800     SH1818     SH1851     SH1858     SH1859    SH1927    SH1832    SH1835    SH1848    SH1849    SH1918    SH1921    SH1929    SH1930    SH1943    USG572;  South Korea:  SG259       SG435       SG537       SG740       SG883       SH0285       SH0663       SH1845       SH1860       SH1853;  Spain:  SG2045       SG210       SG796       SG840       SH1443;  Sweden:  SH1607       SH1608       SH1757       SH1854       SH1874  SH1938 ;  Singapore:  SH1340       Switzerland:  SH0301       SH0553       SH0719       SH1086       SH1103       SH1114;  Taiwan:  SH0738       SH1121       SH1615       SH1755       SH1758       SH1807       SH1813       Turkey:  SG452       SH0251       SH0328       SH1691       SH1779       SH1781       SH1901       United Kingdom:  SH0589      SH1000      SH1759      SH1776      SH1806      USG432."
    }
  ]
}