  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62834",
      "recalling_firm": "Stryker Orthobiologics, Inc.",
      "address_1": "45 Great Valley Pkwy",
      "address_2": "N/A",
      "postal_code": "19355-1302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution-including the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV",
      "recall_number": "Z-0431-2013",
      "product_description": "Styker Orthobiologics Vita Plasma Separator, a component of Vitagel Surgical Hemostat.    Sterile, single use device designed to prepare plasma from the patient's blood using a tabletop centrifuge.",
      "product_quantity": "3,268",
      "reason_for_recall": "When preparing plasma using VitaPrep, it was reported that a small amount of blood collected in the bottom of the centrifuge cup following the centrifugation step.  Firm has since discovered that if blood is present in the Luer connection of the VitaPrep following the blood draw step, this blood could be displaced to the outside of the syringe when the VitaPrep cap is reattached to the syringe.  T",
      "recall_initiation_date": "20120726",
      "center_classification_date": "20121122",
      "termination_date": "20141029",
      "report_date": "20121128",
      "code_info": "Part number 2113-0013, lot numbers T1203002, T1204005, and T1206001."
    }
  ]
}