  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "There is the potential for the vial volume of \"in use\" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded.  Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1",
      "address_2": "PO BOX 6101",
      "product_quantity": "776",
      "code_info": "Dimension Vista(R) 500 - Siemens Material Number 10284473 and  Dimension Vista(R) 1500 - Siemens Material Number 10444801",
      "center_classification_date": "20131202",
      "distribution_pattern": "Worldwide Distribution: USA (nationwide) and iinternationally to Canada.",
      "state": "DE",
      "product_description": "Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1.    Diagnostic device intended to determine a variety of analytes in human body fluids.",
      "report_date": "20131211",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-0430-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65408",
      "termination_date": "20140410",
      "recall_initiation_date": "20130520",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}