  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hazelwood",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63479",
      "recalling_firm": "Biomerieux Inc",
      "address_1": "595 Anglum Rd",
      "address_2": "N/A",
      "postal_code": "63042-2320",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of: AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, PA, RI, TN, TX, VA and WI.  Distribution was also made to Canada, military facilities and other foriegn countries.",
      "recall_number": "Z-0430-2013",
      "product_description": "0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter    Product Usage:  The concentrated sodium chloride is used to prepare inoculum that is analyzed by the VITEK 2 instrument. The VITEK 2 user manual instructs all users to prepare inoculum with 0.45%-0.50% sodium chloride injections, and the system is designed to use 0.45%-0.50% sodium chloride concentration only.",
      "product_quantity": "199 cases/14-1000mL bags",
      "reason_for_recall": "The firm inventoried and distributed 0.9% Sodium Chloride Saline bags as 0.45% Sodium Chloride Saline bags for use with the VITEK 2 system.",
      "recall_initiation_date": "20121017",
      "center_classification_date": "20121121",
      "termination_date": "20131205",
      "report_date": "20121128",
      "code_info": "Lot number C866715, EXP OCT 13"
    }
  ]
}