  "results": [
    {
      "status": "Terminated",
      "city": "Winnipeg",
      "state": "Manitoba",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63500",
      "recalling_firm": "Innovative Magnetic Resonance Imaging Systems Inc.",
      "address_1": "78 Innovation Drive",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution, including the states of FL, GA, IN, MD,MA, MI, MN, MO, TN, UT, and VA.",
      "recall_number": "Z-0428-2013",
      "product_description": "IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.",
      "product_quantity": "23",
      "reason_for_recall": "IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.",
      "recall_initiation_date": "20121019",
      "center_classification_date": "20121121",
      "termination_date": "20121211",
      "report_date": "20121128",
      "code_info": "10002845, 10002341, 10003244, 10002794, 10003243, 10003629, 10003801, 10002654, 10002796, 10002795, 10002846, 10002847, 10003830, 10003064, 10005201,10003754, 10004513"
    }
  ]
}