{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Acton",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91127",
      "recalling_firm": "Insulet Corporation",
      "address_1": "100 Nagog Park",
      "address_2": "N/A",
      "postal_code": "01720-3440",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide.",
      "recall_number": "Z-0423-2023",
      "product_description": "Omnipod 5 Automated Insulin Delivery System  Product catalog numbers:  PT-000409: ASM, Omnipod 5, PDM, Programmed  PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US  PT-000429: Cable, Charging, Omnipod 5, Horizon PDM",
      "product_quantity": "129,086",
      "reason_for_recall": "The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.",
      "recall_initiation_date": "20221114",
      "center_classification_date": "20221207",
      "report_date": "20221214",
      "code_info": "ALL serial numbers/ controller lots are currently in scope of this recall"
    }
  ]
}