  "results": [
    {
      "status": "Terminated",
      "city": "Lower Hutt",
      "state": "",
      "country": "New Zealand",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81166",
      "recalling_firm": "AROA Biosurgery, LTD.",
      "address_1": "69 Gracefield Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "The products were distributed to the following US states:  AL, CA, FL, IN, MA, MI, NH, and NY.",
      "recall_number": "Z-0423-2019",
      "product_description": "OviTex 1S Reinforced BioScaffold 10x12cm, Part Number F10256-1012G",
      "product_quantity": "881 total",
      "reason_for_recall": "Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.",
      "recall_initiation_date": "20180406",
      "center_classification_date": "20181109",
      "termination_date": "20190415",
      "report_date": "20181121",
      "code_info": "ERT-6E05  ERT-6E06  ERT-7K02",
      "more_code_info": ""
    }
  ]
}