{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91088",
      "recalling_firm": "Getinge Usa Sales Inc",
      "address_1": "1 Geoffrey Way",
      "address_2": "N/A",
      "postal_code": "07470-2035",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.",
      "recall_number": "Z-0422-2023",
      "product_description": "Flow-c Anesthesia System (Product Code 6887700)  Flow-e Anesthesia System (Product Code 6887900)",
      "product_quantity": "65 US",
      "reason_for_recall": "The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.",
      "recall_initiation_date": "20221107",
      "center_classification_date": "20221207",
      "report_date": "20221214",
      "code_info": "Flow-c UDI-DI: 7325710010457 Lot/Serial Numbers 4003 4777 4783 5024 5025 5026 5027 5028 5043 5044 5045 5046 5047 5048 5049 5050 5078  Flow-e UDI-DI: 7325710009765 Lot/Serial Numbers 50012 50013 50014 50015 50016 50017 50030 50031 50033 50034 50035 50036 50037 50038 50039 50040 50041 50042 50043 50046 50047 50048 50051 50052 50053 50054 50055 50056 50057 50058 50059 50060 50061 50062 50069 50070 50071 50072 50073 50074 50084 50085 50086 50087 50088 50089 50090 50091 50092 50093 50094 50096 50097 50108 50109 50110 50111 50132"
    }
  ]
}