  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91071",
      "recalling_firm": "Integrity Implants Inc.",
      "address_1": "354 Hiatt Drive",
      "address_2": "N/A",
      "postal_code": "33418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US distribution to California, Indiana, New York, Oklahoma, Washington DC",
      "recall_number": "Z-0421-2023",
      "product_description": "LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545.     Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments.     Used in spinal fixation surgery.",
      "product_quantity": "50 screws",
      "reason_for_recall": "Potential for intraoperative screw head separation at weld location.",
      "recall_initiation_date": "20220803",
      "center_classification_date": "20221205",
      "termination_date": "20231023",
      "report_date": "20221214",
      "code_info": "Part/Model Number LS-N6SG6545; UDI: (01)00818613027815(10)10210067; Lot Number: 10210067"
    }
  ]
}