      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.",
      "address_2": "",
      "product_quantity": "638 units",
      "code_info": "Item No. 814309180, Lot No. 459250, Exp. Date 05/27/2027, UDI (01)00887868037962 (17)270527 (10)459250",
      "center_classification_date": "20180129",
      "distribution_pattern": "US nationwide distribution. Also Portugal and the Netherlands.",
      "state": "IN",
      "product_description": "Affixus Hip Fracture Nail, 125 DEG 9MM X 180MM",
      "report_date": "20180207",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0421-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78763",
      "termination_date": "20190329",
      "more_code_info": "",
      "recall_initiation_date": "20170807",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}