{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Noblesville",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63580",
      "recalling_firm": "King Systems Corp.",
      "address_1": "15011 Herriman Blvd",
      "address_2": "N/A",
      "postal_code": "46060",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC",
      "recall_number": "Z-0419-2013",
      "product_description": "KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile  Size 3 / Child Silicone King LAD¿ Patient Size: 30-50kg     Product Usage:  Laryngeal airway",
      "product_quantity": "263 cases of product. (2,630 devices)",
      "reason_for_recall": "The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change.  The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO",
      "recall_initiation_date": "20121102",
      "center_classification_date": "20121121",
      "termination_date": "20130528",
      "report_date": "20121128",
      "code_info": "REF: LAD-303, Lot #K120618"
    }
  ]
}