  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plymouth",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69632",
      "recalling_firm": "Materialise USA LLC",
      "address_1": "44650 Helm Ct",
      "address_2": "N/A",
      "postal_code": "48170-6061",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Nationwide Distribution to MN only",
      "recall_number": "Z-0418-2015",
      "product_description": "Zimmer Patient Specific Instrument PERSONA CR Pin Guides (Femur and Tibia) is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.",
      "product_quantity": "1",
      "reason_for_recall": "The box of guides for a specific case arrived intact but contained two (2) femoral guides instead of one (1) femoral guide and one (1) tibial guide.  The device did not meet specification: One tibia guide missing and one femur guide contained an edge which was +0.04 mm over specification.",
      "recall_initiation_date": "20141016",
      "center_classification_date": "20141121",
      "termination_date": "20150217",
      "report_date": "20141203",
      "code_info": "00-5970-000-25  JTOMA-JL-0636-L8  Pin Guide LOT#: 56602902",
      "more_code_info": ""
    }
  ]
}