  "results": [
    {
      "status": "Terminated",
      "city": "Beverly",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63075",
      "recalling_firm": "SpineFrontier, Inc.",
      "address_1": "500 Cummings Ctr Ste 3500",
      "address_2": "N/A",
      "postal_code": "01915-6516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide including the states of: MA and TX.",
      "recall_number": "Z-0417-2013",
      "product_description": "S-Lift Implant Guides (also known as S-LIF)  Compressor Part Number: 11-50223      Product Usage:  Surgical Guides allow a surgeon to place implants into tight spaces, such as a vertebral disc space, without having the implant contact the surrounding bone or tissue. In the case of S-LIFT Implant Guides, they are used to place vertebral bodies into the disc space without the body contacting the superior or inferior vertebrae.",
      "product_quantity": "19 units",
      "reason_for_recall": "When using the optional S-LIFT (also known as S-LIF) Implant Guides without stops, there is a known risk of moving the Implant Guides, and possibly the implant, too far into the disc space creating the potential for injury to the patient.  The instructions for Use for the S-Lift Implant Guides will be amended to add the following warning: \"Care is to be taken to not over insert the Implant Guides",
      "recall_initiation_date": "20120822",
      "center_classification_date": "20121121",
      "termination_date": "20150611",
      "report_date": "20121128",
      "code_info": "Lot Numbers: 012064-001"
    }
  ]
}