{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66864",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK",
      "recall_number": "Z-0413-2014",
      "product_description": "SIGMA¿ LCS¿ HIGH PERFORMANCE\"  MBT KEEL PUNCH CEMENTED  Size 2-3      HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty",
      "product_quantity": "29,235 units in total / 2008 through November 2013",
      "reason_for_recall": "DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.",
      "recall_initiation_date": "20131118",
      "center_classification_date": "20131127",
      "termination_date": "20150224",
      "report_date": "20131204",
      "code_info": "Item #950502011"
    }
  ]
}