  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Buffalo Grove",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89009",
      "recalling_firm": "Leica Microsystems, Inc.",
      "address_1": "1700 Leider Ln",
      "address_2": "N/A",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of CA, DC, FL, GA, IL, IN, MA, MI, MD, NY, OH, OR, PA, TN, TX, VA, WI, and Puerto Rico. The countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Burundi, Canada, Chile, China, Colombia, Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, France, Germany, Georgia, Great Britain, Greece, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, South Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal,  North Macedonia, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam.",
      "recall_number": "Z-0412-2022",
      "product_description": "Leica Microsystems M220 F12 Microscope Systems",
      "product_quantity": "2,800",
      "reason_for_recall": "In May 2021, Leica Microsystems received one (1) complaint pertaining to the M220 F12 Surgical Microscope system. There was no patient injury reported for this complaint. However, the identified defect carries some potential injury risk. In the event that this defect may occur, the M220 optics carrier may unintentionally drop into the surgical field, risking contact with the patient.",
      "recall_initiation_date": "20211103",
      "center_classification_date": "20211222",
      "termination_date": "20240618",
      "report_date": "20211229",
      "code_info": "Part #: 10448367  UDI #: 07630003552188   U.S. Serial #'s: 180709002, 191109001, 261016002, 250815001, 280813001, 251111006, 61209001, 260810003, 240815001, 171116001, 190815001, 151209004, 261016001, 70715001, 60713002, 140510004, 180712001, 300910003, 161013001, 270612001, 140510005, 250512002, 10712002, 130913001, 51112001, 290414001, 260311001, 180515001, 270813001, 231014001, 110614001, 210814001, 180515002, 111212001, 270812001, 100614001, 171214002, 290514003, 101212001, 291013001, 50614002, 81012001, 30514001, 101212002, 40611002, 50614003, 171116002, 200314001, 170212001, 30311002, 150211001, 130913002, 60210005, 100212001, 290509001, 111113001, 120615003, 100212002, 300411006, 110315001, 210612002, 100212003, 210612001, 30614001, 20816001, 100212004, 260816001, 260816002, 291113001, 100212005, 21109002, 260811001, 100810004, 90112002  Canada Serial #'s: 260311010, 240414001, 70414002"
    }
  ]
}