  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Beverly",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63144",
      "recalling_firm": "SpineFrontier, Inc.",
      "address_1": "500 Cummings Ctr Ste 3500",
      "address_2": "N/A",
      "postal_code": "01915-6516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including PR and the states of FL, MA, MO, OK, and TX, and the country of Mexico.",
      "recall_number": "Z-0412-2013",
      "product_description": "Spine Frontier Inspan compressors  Part Number: 11-60004.    Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone, resulting in fixation of the implant.",
      "product_quantity": "17 units",
      "reason_for_recall": "Inspan Compressors may break at weld causing a loss of compression",
      "recall_initiation_date": "20120828",
      "center_classification_date": "20121121",
      "termination_date": "20150611",
      "report_date": "20121128",
      "code_info": "Lot Numbers: SFP02323, SFP02323R, and SFP02323RR"
    }
  ]
}