      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.",
      "address_2": "",
      "product_quantity": "4,207 units total",
      "code_info": "Model Number, UDI Number, (Serial Number):    2097674-001-US, Not Applicable, (5470345-30);  2097674-001-US, Not Applicable, (5470345-31);  2097674-001-US, Not Applicable, (5470345-32);    1011-9000-000, Not Applicable, (ANAK00774);  1011-9000-000, Not Applicable, (ANAK00775);  1011-9000-000, Not Applicable, (ANAL01201);  1011-9000-000, Not Applicable, (ANAL01234);  1011-9000-000, Not Applicable, (ANAL01235);  1011-9000-000, Not Applicable, (ANAL01236);    1011-9000-000, Not Applicable, (ANAL01237);  1011-9000-000, Not Applicable, (ANAL01238);  1011-9000-000, Not Applicable, (ANAL01240);    1011-9000-000, Not Applicable, (ANAL01494);  1011-9000-000, Not Applicable, (ANAL01495);  1011-9000-000, Not Applicable, (ANAL01551);    1011-9000-000, Not Applicable, (ANAM01023);    1011-9000-000, Not Applicable, (ANAM01024);    1011-9000-000, Not Applicable, (ANAM01025);    1011-9000-000, Not Applicable, (ANAM01026);    1011-9000-000, Not Applicable, (ANAM01027);    1011-9000-000, Not Applicable, (ANAM01104);    1011-9000-000, Not Applicable, (ANAM01105);    1011-9000-000, Not Applicable, (ANAP00438);    1011-9000-000, Not Applicable, (ANAP00439);    1011-9000-000, Not Applicable, (ANAQ01188);    1011-9000-000, Not Applicable, (ANAR05264);    1011-9000-000, Not Applicable, (ANAR05265);    1011-9000-000, Not Applicable, (ANAR05610);    1011-9000-000, Not Applicable, (ANAR05611);    1011-9000-000, Not Applicable, (ANAS00202);    1011-9000-000, Not Applicable, (ANAS00203);    1011-9000-000, Not Applicable, (ANAS00204);    2097674-001-US, Not Applicable, (5479630-1);    ANAK00798, Not Applicable (1011-9000-000);    ANAK00799, Not Applicable (1011-9000-000);    ANAK00803, Not Applicable (1011-9000-000);    ANAK00807, Not Applicable (1011-9000-000);    ANAK00814, Not Applicable (1011-9000-000);    ANAK00839, Not Applicable (1011-9000-000);    ANAK00864, Not Applicable (1011-9000-000);    5481223-1, Not Applicable (2097674-001-US);    5481223-10, Not Applicable (2097674-001-US);    5481223-11, Not Applicable (2097674-001-US);    5481223-12, Not Applicable (2097674-001-US);    5481223-2, Not Applicable (2097674-001-US);    5481223-3, Not Applicable (2097674-001-US);    5481223-4, Not Applicable (2097674-001-US);    5481223-5, Not Applicable (2097674-001-US);    5481223-6, Not Applicable (2097674-001-US);    5481223-7, Not Applicable (2097674-001-US);    5481223-8, Not Applicable (2097674-001-US);    5481223-9, Not Applicable (2097674-001-US);    ANAP00712, Not Applicable (1011-9000-000);    ANAP00713, Not Applicable (1011-9000-000);    ANAP00714, Not Applicable (1011-9000-000);    ANAN01125, Not Applicable (1011-9000-000);    ANAN01138, Not Applicable (1011-9000-000);    4652664-1, Not Applicable (2097674-001-US);    4652664-10, Not Applicable (2097674-001-US);    4652664-11, Not Applicable (2097674-001-US);    4652664-12, Not Applicable (2097674-001-US);    4652664-13, Not Applicable (2097674-001-US);    4652664-2, Not Applicable (2097674-001-US);    4652664-3, Not Applicable (2097674-001-US);    4652664-4, Not Applicable (2097674-001-US);    4652664-5, Not Applicable (2097674-001-US);    4652664-6, Not Applicable (2097674-001-US);    4652664-7, Not Applicable (2097674-001-US);    4652664-8, Not Applicable (2097674-001-US);    4652664-9, Not Applicable (2097674-001-US);    ANAK00156, Not Applicable (1011-9000-000);    ANAK00157, Not Applicable (1011-9000-000);    ANAK00158, Not Applicable (1011-9000-000);    ANAK00159, Not Applicable (1011-9000-000);    ANAK00511, Not Applicable (1011-9000-000);    ANAK00512, Not Applicable (1011-9000-000);    ANAK00513, Not Applicable (1011-9000-000);    ANAK00514, Not Applicable (1011-9000-000);    ANAN01051, Not Applicable (1011-9000-000);    ANAN01052, Not Applicable (1011-9000-000);    ANAP00545, Not Applicable (1011-9000-000);    ANAP00546, Not Applicable (1011-9000-000);    ANAP00708, Not Applicable (1011-9000-000);    ANAL01400, Not Applicable (1011-9000-000);    ANAL01401, Not Applicable (1011-9000-000);    ANAM00449, Not Applicable (1011-9000-000);    ANAM01089, Not Applicable (1011-9000-000);    5453450-1, Not Applicable (2097674-001);    5453450-2, Not Applicable (2097674-001);    5453450-3, Not Applicable (2097674-001);    5453450-4, Not Applicable (2097674-001);    5453450-5, Not Applicable (2097674-001);    5453450-6, Not Applicable (2097674-001);    5453450-7, Not Applicable (2097674-001);    5453450-8, Not Applicable (2097674-001);    5480532-1, Not Applicable (2097674-001);    5447751-1, Not Applicable (2097674-001);    5447751-10, Not Applicable (2097674-001);    5447751-11, Not Applicable (2097674-001);    5447751-12, Not Applicable (2097674-001);    5447751-13, Not Applicable (2097674-001);    5447751-14, Not Applicable (2097674-001);    5447751-15, Not Applicable (2097674-001);    5447751-16, Not Applicable (2097674-001);    5447751-17, Not Applicable (2097674-001);    5447751-18, Not Applicable (2097674-001);    5447751-19, Not Applicable (2097674-001);    5447751-2, Not Applicable (2097674-001);    5447751-20, Not Applicable (2097674-001);    5447751-21, Not Applicable (2097674-001);    5447751-22, Not Applicable (2097674-001);    5447751-23, Not Applicable (2097674-001);    5447751-24, Not Applicable (2097674-001);    5447751-3, Not Applicable (2097674-001);    5447751-4, Not Applicable (2097674-001);    5447751-5, Not Applicable (2097674-001);    5447751-6, Not Applicable (2097674-001);    5447751-7, Not Applicable (2097674-001);    5447751-8, Not Applicable (2097674-001);    5447751-9, Not Applicable (2097674-001);    5452649-1, Not Applicable (2097674-001);    5452649-2, Not Applicable (2097674-001);    5474323-1, Not Applicable (2097674-DE);    ANAR05021, Not Applicable (1011-9000-000);    5473625-5, Not Applicable (2097674-NL);    5473625-6, Not Applicable (2097674-NL);    5473625-7, Not Applicable (2097674-NL);    ANAK00330, Not Applicable (1011-9000-000);    ANAK00332, Not Applicable (1011-9000-000);    ANAK00334, Not Applicable (1011-9000-000);    ANAL00806, Not Applicable (1011-9000-000);    ANAL00807, Not Applicable (1011-9000-000);    ANAL00808, Not Applicable (1011-9000-000);    ANAL00809, Not Applicable (1011-9000-000);    ANAN01010, Not Applicable (1011-9000-000);    ANAN01011, Not Applicable (1011-9000-000);    ANAN01012, Not Applicable (1011-9000-000);    ANAN01013, Not Applicable (1011-9000-000);    ANAN01014, Not Applicable (1011-9000-000);    ANAP00399, Not Applicable (1011-9000-000);    ANAP01133, Not Applicable (1011-9000-000);    ANAR00344, Not Applicable (1011-9000-000);    ANAR00896, Not Applicable (1011-9000-000);    ANAR05978, Not Applicable (1011-9000-000);    ANAR05979, Not Applicable (1011-9000-000);    ANAR05980, Not Applicable (1011-9000-000);    ANAR05981, Not Applicable (1011-9000-000);    ANAR05982, Not Applicable (1011-9000-000);    ANAR05983, Not Applicable (1011-9000-000);    ANAR05984, Not Applicable (1011-9000-000);    ANAS00709, Not Applicable (1011-9000-000);    ANAS00710, Not Applicable (1011-9000-000);    ANAS00711, Not Applicable (1011-9000-000);    5472401-2, Not Applicable (2097674-001);    ANAP00132, Not Applicable (1011-9000-000);    5472557-4, Not Applicable (2097674-001);    5472557-5, Not Applicable (2097674-001);    5472557-6, Not Applicable (2097674-001);    ANAN01045, Not Applicable (1011-9000-000);    ANAN01046, Not Applicable (1011-9000-000);    ANAR00327, Not Applicable (1011-9000-000);    ANAS00345, Not Applicable (1011-9000-000);    5454764-3, Not Applicable (2097674-001);    ANAP00584, Not Applicable (1011-9000-000);    ANAP00585, Not Applicable (1011-9000-000);    ANAP00586, Not Applicable (1011-9000-000);    ANAP00587, Not Applicable (1011-9000-000);    ANAP00588, Not Applicable (1011-9000-000);    ANAP00589, Not Applicable (1011-9000-000);    ANAP00825, Not Applicable (1011-9000-000);    ANAP01215, Not Applicable (1011-9000-000);    ANAS00738, Not Applicable (1011-9000-000);    ANAL01390, Not Applicable (1011-9000-000);    ANAL01391, Not Applicable (1011-9000-000);    ANAS00391, Not Applicable (1011-9000-000);    ANAR00492, Not Applicable (1011-9000-000);    ANAR00493, Not Applicable (1011-9000-000);    ANAR00494, Not Applicable (1011-9000-000);    ANAR00495, Not Applicable (1011-9000-000);    ANAR00496, Not Applicable (1011-9000-000);    ANAR00497, Not Applicable (1011-9000-000);    ANAR00498, Not Applicable (1011-9000-000);    ANAR00499, Not Applicable (1011-9000-000);    ANAR00500, Not Applicable (1011-9000-000);    ANAR00501, Not Applicable (1011-9000-000);    ANAR00756, Not Applicable (1011-9000-000);    ANAR00757, Not Applicable (1011-9000-000);    ANAL00705, Not Applicable (1011-9000-000);    ANAL01040, Not Applicable (1011-9000-000);    ANAL01041, Not Applicable (1011-9000-000);    ANAL01042, Not Applicable (1011-9000-000);    ANAL01043, Not Applicable (1011-9000-000);    ANAL01044, Not Applicable (1011-9000-000);    ANAL01045, Not Applicable (1011-9000-000);    ANAL01046, Not Applicable (1011-9000-000);    ANAL01047, Not Applicable (1011-9000-000);    ANAL01048, Not Applicable (1011-9000-000);    ANAL01050, Not Applicable (1011-9000-000);    ANAM00335, Not Applicable (1011-9000-000);    ANAL00192, Not Applicable (1011-9000-000);    ANAL00754, Not Applicable (1011-9000-000);    ANAL00755, Not Applicable (1011-9000-000);    ANAL00756, Not Applicable (1011-9000-000);    ANAL00757, Not Applicable (1011-9000-000);    ANAL00764, Not Applicable (1011-9000-000);    ANAL00769, Not Applicable (1011-9000-000);    ANAL00776, Not Applicable (1011-9000-000);    ANAL00777, Not Applicable (1011-9000-000);    ANAL00778, Not Applicable (1011-9000-000);    ANAL00779, Not Applicable (1011-9000-000);    ANAL00780, Not Applicable (1011-9000-000);    ANAL00781, Not Applicable (1011-9000-000);    ANAL00782, Not Applicable (1011-9000-000);    ANAL00783, Not Applicable (1011-9000-000);    ANAL00784, Not Applicable (1011-9000-000);    ANAL00785, Not Applicable (1011-9000-000);    ANAL00786, Not Applicable (1011-9000-000);    ANAL00787, Not Applicable (1011-9000-000);    ANAL00788, Not Applicable (1011-9000-000);    ANAL00789, Not Applicable (1011-9000-000);    ANAL00790, Not Applicable (1011-9000-000);    ANAL00791, Not Applicable (1011-9000-000);    ANAL00792, Not Applicable (1011-9000-000);    ANAL01541, Not Applicable (1011-9000-000);    ANAL01542, Not Applicable (1011-9000-000);    ANAP00810, Not Applicable (1011-9000-000);    ANAP00811, Not Applicable (1011-9000-000);    ANAS00938, Not Applicable (1011-9000-000);    ANAP01070, Not Applicable (1011-9000-000);    5473409-1, Not Applicable (2097674-FR);    5473409-2, Not Applicable (2097674-FR);    5473409-3, Not Applicable (2097674-FR);    ANAP00517, Not Applicable (1011-9000-000);    ANAP00518, Not Applicable (1011-9000-000);    ANAP00519, Not Applicable (1011-9000-000);    ANAM01291, Not Applicable (1011-9000-000);    ANAM01292, Not Applicable (1011-9000-000);    ANAM01293, Not Applicable (1011-9000-000);    ANAN00430, Not Applicable (1011-9000-000);    ANAN00431, Not Applicable (1011-9000-000);    ANAN00432, Not Applicable (1011-9000-000);    ANAN00436, Not Applicable (1011-9000-000);    ANAN00437, Not Applicable (1011-9000-000);    ANAP00300, Not Applicable (1011-9000-000);    ANAQ00766, Not Applicable (1011-9000-000);    ANAS00753, Not Applicable (1011-9000-000);    ANAP00492, Not Applicable (1011-9000-000);    ANAP00493, Not Applicable (1011-9000-000);    ANAM00628, Not Applicable (1011-9000-000);    ANAL01029, Not Applicable (1011-9000-000);    5479515-1, Not Applicable (2097674-JA);    5479515-2, Not Applicable (2097674-JA);    5479515-3, Not Applicable (2097674-JA);    5479515-4, Not Applicable (2097674-JA);    5479515-5, Not Applicable (2097674-JA);    5479515-6, Not Applicable (2097674-JA);    5479515-7, Not Applicable (2097674-JA);    5479515-8, Not Applicable (2097674-JA);    5479515-9, Not Applicable (2097674-JA);    5480673-1, Not Applicable (2097674-JA);    5480673-2, Not Applicable (2097674-JA);    5480673-3, Not Applicable (2097674-JA);    5480673-4, Not Applicable (2097674-JA);    5480673-5, Not Applicable (2097674-JA);    5480673-6, Not Applicable (2097674-JA);    5480673-7, Not Applicable (2097674-JA);    5480673-8, Not Applicable (2097674-JA);    5483069-1, Not Applicable (2097674-JA);    ANAQ01435, Not Applicable (1011-9000-000);    ANAM01312, Not Applicable (1011-9000-000);    ANAM01313, Not Applicable (1011-9000-000);    ANAM01315, Not Applicable (1011-9000-000);    ANAM01317, Not Applicable (1011-9000-000);    ANAM01318, Not Applicable (1011-9000-000);    ANAM01320, Not Applicable (1011-9000-000);    ANAM01324, Not Applicable (1011-9000-000);    ANAM01325, Not Applicable (1011-9000-000);    ANAL01313, Not Applicable (1011-9000-000);    ANAL01314, Not Applicable (1011-9000-000);    ANAL01315, Not Applicable (1011-9000-000);    ANAL01316, Not Applicable (1011-9000-000);    ANAL01317, Not Applicable (1011-9000-000);    ANAL01318, Not Applicable (1011-9000-000);    ANAL01319, Not Applicable (1011-9000-000);    ANAL01320, Not Applicable (1011-9000-000);    ANAM01302, Not Applicable (1011-9000-000);    ANAM01304, Not Applicable (1011-9000-000);    ANAN00487, Not Applicable (1011-9000-000);    ANAN00488, Not Applicable (1011-9000-000);    ANAN00489, Not Applicable (1011-9000-000);    ANAN00490, Not Applicable (1011-9000-000);    ANAN00491, Not Applicable (1011-9000-000);    ANAN00492, Not Applicable (1011-9000-000);    ANAR05143, Not Applicable (1011-9000-000);    ANAR05144, Not Applicable (1011-9000-000);    5450447-2, Not Applicable (2097674-NL);    5450447-4, Not Applicable (2097674-NL);    5454351-1, Not Applicable (2097674-NL);    5454351-2, Not Applicable (2097674-NL);    ANAR00198, Not Applicable (1011-9000-000);    ANAR00199, Not Applicable (1011-9000-000);    ANAR00200, Not Applicable (1011-9000-000);    ANAR00279, Not Applicable (1011-9000-000);    ANAM01311, Not Applicable (1011-9000-000);    ANAM01314, Not Applicable (1011-9000-000);    ANAM01316, Not Applicable (1011-9000-000);    ANAM01319, Not Applicable (1011-9000-000);    ANAM01321, Not Applicable (1011-9000-000);    ANAM01322, Not Applicable (1011-9000-000);    ANAM01323, Not Applicable (1011-9000-000);    5456012-1, Not Applicable (2097674-001);    5456012-2, Not Applicable (2097674-001);    5456012-3, Not Applicable (2097674-001);    5456719-1, Not Applicable (2097674-001);    ANAN01065, Not Applicable (1011-9000-000);    ANAN01066, Not Applicable (1011-9000-000);    5476468-1, Not Applicable (2097674-001);    5476468-10, Not Applicable (2097674-001);    5476468-11, Not Applicable (2097674-001);    5476468-12, Not Applicable (2097674-001);    5476468-13, Not Applicable (2097674-001);    5476468-14, Not Applicable (2097674-001);    5476468-15, Not Applicable (2097674-001);    5476468-16, Not Applicable (2097674-001);    5476468-17, Not Applicable (2097674-001);    5476468-2, Not Applicable (2097674-001);    5476468-3, Not Applicable (2097674-001);    5476468-4, Not Applicable (2097674-001);    5476468-5, Not Applicable (2097674-001);    5476468-6, Not Applicable (2097674-001);    5476468-7, Not Applicable (2097674-001);    5476468-8, Not Applicable (2097674-001);    5476468-9, Not Applicable (2097674-001);      Additional Serial Numbers - Model Number, UDI Number, (Serial Number):  2097674-001-US, Not Applicable, (APWV01271);   2097674-001-US, Not Applicable, (APWU01812);   2097674-001-US, Not Applicable, (ANAR00714);   2097674-001-US, Not Applicable, (ANAR00741);   2097674-001-US, Not Applicable, (ANAS00330);   2097674-001-US, Not Applicable, (ANAS01319);   2097674-001-US, Not Applicable, (ANAS01320);   2097674-001-US, Not Applicable, (APWU00486);   2097674-001-US, Not Applicable, (APWU00487);   2097674-001-US, Not Applicable, (APWU00575);   2097674-001-US, Not Applicable, (APWU01267);   2097674-001-US, Not Applicable, (APWX00154);   2097674-001-US, Not Applicable, (APWX00155);   2097674-001-US, Not Applicable, (APWX00156);   2097674-001-US, Not Applicable, (APWX00157);   1011-9000-000, Not Applicable, (ANAS00456);   2097674-001-US, Not Applicable, (ANAM00764);   2097674-001-US, Not Applicable, (ANAM00765);   2097674-001-US, Not Applicable, (ANAM00766);   2097674-001-US, Not Applicable, (ANAM00767);   2097674-001-US, Not Applicable, (ANAM00768);   2097674-001-US, Not Applicable, (ANAM00769);   2097674-001-US, Not Applicable, (ANAM00770);   2097674-001-US, Not Applicable, (ANAM00771);   2097674-001-US, Not Applicable, (ANAM00772);   2097674-001-US, Not Applicable, (ANAM00773);   2097674-001-US, Not Applicable, (ANAM00774);   2097674-001-US, Not Applicable, (ANAM00775);   2097674-001-US, Not Applicable, (APWV00946);   2097674-001, Not Applicable, (ANAL00264);   2097674-001, Not Applicable, (ANAL00265);   2097674-001, Not Applicable, (ANAL00266);   2097674-001, Not Applicable, (ANAK00415);   2097674-001, Not Applicable, (ANAK00703);   2097674-001, Not Applicable, (ANAK00705);   2097674-001, Not Applicable, (ANAK00707);   2097674-001, Not Applicable, (ANAK00709);   2097674-001, Not Applicable, (ANAK00410);   2097674-001, Not Applicable, (ANAK00411);   2097674-001, Not Applicable, (ANAK00413);   2097674-001, Not Applicable, (ANAK00414);   2097674-001, Not Applicable, (ANAK00416);   2097674-001, Not Applicable, (ANAK00417);   2097674-001, Not Applicable, (ANAK00421);   2097674-001, Not Applicable, (ANAL00893);   2097674-001, Not Applicable, (APWU02246);   2097674-001, Not Applicable, (APWV00404);   2097674-001, Not Applicable, (ANAK00412);   2097674-001, Not Applicable, (ANAK00418);   2097674-001, Not Applicable, (ANAK00419);   2097674-001, Not Applicable, (ANAK00420);   2097674-001, Not Applicable, (ANAK00704);   2097674-001, Not Applicable, (ANAK00706);   2097674-001, Not Applicable, (ANAK00708);   2097674-001, Not Applicable, (APWW00334);   1011-9000-000, Not Applicable, (ANAR05371);   2097674-DE, Not Applicable, (ANAS00870);   1011-9000-000, Not Applicable, (ANAR00333);   1011-9000-000, Not Applicable, (ANAJ00122);   1011-9000-000, Not Applicable, (ANAQ01270);   1011-9000-000, Not Applicable, (ANAQ01271);   1011-9000-000, Not Applicable, (ANAQ01272);   1011-9000-000, Not Applicable, (ANAQ01273);   1011-9000-000, Not Applicable, (ANAQ01274);   1011-9000-000, Not Applicable, (ANAQ01275);   1011-9000-000, Not Applicable, (ANAQ01276);   1011-9000-000, Not Applicable, (ANAQ01277);   1011-9000-000, Not Applicable, (ANAR00187);   1011-9000-000, Not Applicable, (ANAR00188);   1011-9000-000, Not Applicable, (ANAR00189);   1011-9000-000, Not Applicable, (ANAR00190);   1011-9000-000, Not Applicable, (ANAR00191);   1011-9000-000, Not Applicable, (ANAR00192);   1011-9000-000, Not Applicable, (ANAR00334);   1011-9000-000, Not Applicable, (ANAR00701);   1011-9000-000, Not Applicable, (APWU02039);   1011-9000-000, Not Applicable, (APWW01068);   2097674-001, Not Applicable, (ANAQ00825);   1011-9000-000, Not Applicable, (ANAR05370);   2097674-FR, Not Applicable, (ANAN01100);   1011-9000-000, Not Applicable, (ANAT00157);   1011-9000-000, Not Applicable, (ANAT00158);   1011-9000-000, Not Applicable, (ANAT00156);   1011-9000-000, Not Applicable, (ANAR05372);",
      "center_classification_date": "20181109",
      "distribution_pattern": "Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.",
      "state": "WI",
      "product_description": "Aisys Anesthesia System.  Sold under the following product names:  AISYS, AISYS 10.X TO 11.X UPGRADE KIT DUTCH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH, AISYS 10.X TO 11.X UPGRADE KIT ENGLISH FOR USA, AISYS 10.X TO 11.X UPGRADE KIT FRENCH, AISYS 10.X TO 11.X UPGRADE KIT GERMAN, AISYS 10.X TO 11.X UPGRADE KIT JAPANESE.",
      "report_date": "20181121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-0411-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81182",
      "termination_date": "20210412",
      "more_code_info": "",
      "recall_initiation_date": "20180927",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}