  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Spencer",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72364",
      "recalling_firm": "Boston Scientific Corp",
      "address_1": "780 Brookside Dr",
      "address_2": "N/A",
      "postal_code": "47460-1080",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : IN, NV, CA, PA, CT, TN, NY, WI, and WA, and Internationally to Ireland.",
      "recall_number": "Z-0402-2016",
      "product_description": "One Step Button, Low Profile Initial Placement PEG Kit, gastro-enterostomy tube.",
      "product_quantity": "22 units",
      "reason_for_recall": "Boston Scientific Corp. initiated a voluntary recall of specific lots of the EndoVive One Step Button, including the Decompression Tube and Low Profile Replacement Button devices because they may contain decompression tube adapters with a protrusion that could impede its ability to be introduced or removed from the PEG Button.",
      "recall_initiation_date": "20150911",
      "center_classification_date": "20151210",
      "termination_date": "20160517",
      "report_date": "20151216",
      "code_info": "Material Number (UPN) M00563020, Lot # 18184947, Exp. April 30, 2017; Material Number  M00563080, Lot # 18179867, Exp.  April 30, 2017; & Lot # 18212410 , Exp. May 31, 2017; Material Number M00568510, Lot # 18296155, Exp. May 31, 2018.",
      "more_code_info": ""
    }
  ]
}