  "results": [
    {
      "status": "Terminated",
      "city": "Fajardo",
      "state": "PR",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69448",
      "recalling_firm": "Customed, Inc",
      "address_1": "Calle Igualdad Final #7",
      "address_2": "N/A",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to Florida, New York and Puerto Rico..",
      "recall_number": "Z-0402-2015",
      "product_description": "Femoral Angiography Pack, Catalog number 900-796.    Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.",
      "product_quantity": "236",
      "reason_for_recall": "Significant compromised sterility, not limited to potentially damaged packaging.  Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.",
      "recall_initiation_date": "20141008",
      "center_classification_date": "20141128",
      "termination_date": "20170814",
      "report_date": "20141210",
      "code_info": "Lot numbers  140513145  140513340  140613848  140814683  140915202"
    }
  ]
}