  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66815",
      "recalling_firm": "Advanced Sterilization Products",
      "address_1": "33 Technology Dr",
      "address_2": "N/A",
      "postal_code": "92618-2346",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0402-2014",
      "product_description": "STERRAD 100NX, Product Code: 10104    The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.",
      "product_quantity": "Total 4,381 units",
      "reason_for_recall": "Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.",
      "recall_initiation_date": "20131111",
      "center_classification_date": "20131126",
      "termination_date": "20140122",
      "report_date": "20131204",
      "code_info": "Please refer to consignee list"
    }
  ]
}