  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Broomfield",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66836",
      "recalling_firm": "Lanx, Inc.",
      "address_1": "310 Interlocken Pkwy Ste 120",
      "address_2": "N/A",
      "postal_code": "80021-3464",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide Distribution including AZ, PA, WI, NY, TX, PR, and TN.",
      "recall_number": "Z-0400-2014",
      "product_description": "LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short (\"Driver\"). Model Number 8630-0107.    The Timberline MPF device is an intervertebral body fusion device consisting of a PEEK intervertebral spacer, titanium plate, and screws. The PEEK spacer has a generally rounded shape with various heights and footprints and has a hollowed out central area to accommodate autogenous bone graft. The upper and lower surfaces have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The titanium plate has holes for receiving bone screws and a central hole for receiving a cover plate to prevent screw back-out. The Timberline MPF system is available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The Timberline MPF system is provided nonsterile. The Driver is used with the Timberline MPF system to drive titanium screws through the plate and into the vertebrae thus locking plate to the vertebrae.",
      "product_quantity": "25 units",
      "reason_for_recall": "Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short (\"Driver\") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.",
      "recall_initiation_date": "20131104",
      "center_classification_date": "20131126",
      "termination_date": "20151229",
      "report_date": "20131204",
      "code_info": "Lot No: TU00090, TU00157 and TU00242"
    }
  ]
}