  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75366",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide.",
      "recall_number": "Z-0396-2017",
      "product_description": "COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test  Catalog number 04528123160",
      "product_quantity": "4976",
      "reason_for_recall": ".Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual  Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles.  This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected.  The issue can lead to erroneously high HbA1c results.",
      "recall_initiation_date": "20161004",
      "center_classification_date": "20161111",
      "termination_date": "20170605",
      "report_date": "20161123",
      "code_info": "Not applicable"
    }
  ]
}