  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66551",
      "recalling_firm": "Synthes USA HQ, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwided Distribution: USA (nationwide) including IL, MO, TX, WA and Internationally to Canada and Switzerland.",
      "recall_number": "Z-0393-2014",
      "product_description": "Synthes Material Mix (Adjustable Cervical Distractor-Right, Matrix Distractor Rack, T-PAL Spacer Remover, Implant Inserter)    Synthes Material Mix: Adjustable Cervical Distractor-Right is used to distract spine segment. Matrix Distractor Rack is used to distract the interbody disc space. T-PAL Spacer Remover is used to remove T-PAL spacer. Implant Inserter is used to insert the ProDisc implant.",
      "product_quantity": "approximately 44",
      "reason_for_recall": "One material batch used for production of the affected parts of the Synthes Material Mix were mixed in with an incorrect material with a lower than specified hardness and corrosion resistance.",
      "recall_initiation_date": "20131015",
      "center_classification_date": "20131126",
      "termination_date": "20150831",
      "report_date": "20131204",
      "code_info": "part no. 396.396 (lot nos. T989765 and T989780), part no. 03.632.087 (lotT987270), part no. 03.812.005 (lot nos. T987734, T988931, and T987623), and part no. 03.820.129 (lot no. T988206)."
    }
  ]
}