  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90967",
      "recalling_firm": "Carefusion 2200 Inc",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines, Saudi Arabia, South Africa, South Korea, Thailand",
      "recall_number": "Z-0391-2023",
      "product_description": "Carefusion V. Mueller\" CUSHING BAYONET BIPOLAR FORCEPS, 2.0MM TIP, INSULATED OVERALL LENGTH 7-1/2\" (190MM), REF F-2004",
      "product_quantity": "1810 units",
      "reason_for_recall": "BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller\" Bipolar Forcep IFU.",
      "recall_initiation_date": "20221102",
      "center_classification_date": "20221203",
      "report_date": "20221214",
      "code_info": "UDI/DI 10885403040917, ALL LOTS"
    }
  ]
}