  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66416",
      "recalling_firm": "Synthes USA HQ, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including GA, FL, CA, UT, IA, AZ, OH and Internationally to Canada.",
      "recall_number": "Z-0391-2014",
      "product_description": "Synthes Matrix 5.5.mm Left and Right Coronal Bender     Product Usage:  The Matrix 5.5.mm Left and Right Coronal Bender are used together to bend the 5.5 Matrix Rods in the coronal plane, and is intended for stabilization of the thoracic, lumbar, sacral, and/or iliac spine through screw and/or hook and rod fixation.",
      "product_quantity": "209",
      "reason_for_recall": "A complaint was received of the tips of the Matrix 5.5.mm Left and Right Coronal Bender breaking while bending a spinal rod component of the Matrix System.",
      "recall_initiation_date": "20130622",
      "center_classification_date": "20131125",
      "termination_date": "20150910",
      "report_date": "20131204",
      "code_info": "Parts 03.632.040 and 03.632.041 with various lot nos.    Item Lot  Number Number     03.632.040  T935338  03.632.040  T939100  03.632.040  T945062  03.632.040  T953982  03.632.040  T956310  03.632.040  T959990  03.632.040  T959995        Item Lot  Number Number     03.632.041  T935343  03.632.041  T939105  03.632.041  T939755  03.632.041  T945067  03.632.041  T952750  03.632.041  T956315  03.632.041  T960000  03.632.041  T960005"
    }
  ]
}