{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66666",
      "recalling_firm": "Synthes USA HQ, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Canada.",
      "recall_number": "Z-0390-2014",
      "product_description": "3.7mm and 5.0mm Dynamic Locking Screw (DLS)      Product Usage:  3.7mm and 5.0mm Dynamic Locking Screw (DLS) in combination with Synthes Locking Compression Plates (LCP) are intended for use in long bone fractures, the fixation of osteopenic bone, the fixation of osteotomies, and for the fixation of non-unions and malunions.",
      "product_quantity": "40248",
      "reason_for_recall": "All lots of 3.7mm and 5.0mm Dynamic Locking Screw (DLS)  were recalled due to complaints of breakages at the distal tip of the pin identified after successful healing and during planned removal of the device.",
      "recall_initiation_date": "20130722",
      "center_classification_date": "20131125",
      "termination_date": "20160818",
      "report_date": "20131204",
      "code_info": "All lots of parts 09.213.022S - 09.213.070S, and 09.223.032S- 09.223.090S."
    }
  ]
}