      "status": "Terminated",
      "city": "Veronaverona",
      "state": "N/A",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84004",
      "recalling_firm": "Orthofix Srl",
      "address_1": "Via Delle Nazioni 9 O 73",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico   OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel",
      "recall_number": "Z-0389-2020",
      "product_description": "ORTHOFIX Catalogue Number: ref: 99-91647UE, UE XCALIBER ANKLE COMPLETE KIT, RX ONLY, UDI: (01)18032937169719",
      "product_quantity": "N/A",
      "reason_for_recall": "As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design.  This could result in lower cutting performance.",
      "recall_initiation_date": "20180927",
      "center_classification_date": "20191114",
      "termination_date": "20220201",
      "report_date": "20191120",
      "code_info": "Lot # B1206515, B1206561, B1209206, B1211406, B1212677, B1215964, B1233839"
    }
  ]
}