  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66542",
      "recalling_firm": "Lin-Zhi International Inc",
      "address_1": "670 Almanor Ave",
      "address_2": "N/A",
      "postal_code": "94085-3513",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution in CA.",
      "recall_number": "Z-0386-2014",
      "product_description": "Single Analyte Urine DAU Control (5 m L  Norbuprenorphine, Level II control  For In Vitro Diagnostic Use Only. store at 2 - 8 degrees C.  Contents: Contains 13 mg/Ml of Norbuprenorphine in human urine with sodium azide as a preservative.    Manufacturer:  Lin-Zhi International, Inc.  670 Almanor Ave.  Sunnyvale, CA 94085    The Norbuprenorphine Drug of Abuse (DUB) Positive control is used as assayed quality control material to monitor the precision of the Lin-Zhi International, Inc. Buprenorphine Enzyme immunoassay on the Synchron clinical analyzer. It in intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine.",
      "product_quantity": "60 items",
      "reason_for_recall": "Product catalog # A68825 Norbuprenorphine 13 ng/mL Level 2 Controls gives a decreased concentration reading due to the degradation of the NBUP analyte  in solution.",
      "recall_initiation_date": "20131003",
      "center_classification_date": "20131125",
      "termination_date": "20131226",
      "report_date": "20131204",
      "code_info": "Catalog #A68825, Lot #1307086; Exp. 3/28/2014/"
    }
  ]
}