{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Research Triangle Park",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75554",
      "recalling_firm": "Teleflex Medical",
      "address_1": "2917 Weck Dr",
      "address_2": "N/A",
      "postal_code": "27709-0186",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.",
      "recall_number": "Z-0383-2017",
      "product_description": "FLEXI NOZZLE SHORT    Product Usage:  These products are used for the delivery of topical anesthesia via an atomized spray to the oral, nasal, pharyngeal or laryngeal mucosa. Teleflex Medical is recalling these products as they may produce a straight stream instead of a fully atomized plume of medication. It is unlikely that serious adverse health consequences will occur in the event of a failure to deliver an atomized plume; however, this may result in inadequate topical anesthesia which may lead to some discomfort, further attempts to deliver topical anesthesia, or the use of alternative methods of anesthesia.",
      "product_quantity": "2,674,650 units in total",
      "reason_for_recall": "Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.",
      "recall_initiation_date": "20161027",
      "center_classification_date": "20161222",
      "termination_date": "20200925",
      "report_date": "20161228",
      "code_info": "Product Code: MAD510L, Lot# 160118, 160324, 160509, 160709, 160810 and 160833."
    }
  ]
}