  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Veronaverona",
      "state": "N/A",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84004",
      "recalling_firm": "Orthofix Srl",
      "address_1": "Via Delle Nazioni 9 O 73",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico   OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel",
      "recall_number": "Z-0382-2020",
      "product_description": "ORTHOFIX Catalogue Number: ref 1-1100301,DRILL BIT Diameter 3.2 MM L. 200 MM TIN COATED- QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856075",
      "product_quantity": "N/A",
      "reason_for_recall": "As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design.  This could result in lower cutting performance.",
      "recall_initiation_date": "20180927",
      "center_classification_date": "20191114",
      "termination_date": "20220201",
      "report_date": "20191120",
      "code_info": "Lot # B1203608, B1210965, B1211752, B1214172,B1218793, 300241,300240,273169"
    }
  ]
}