  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78900",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.",
      "recall_number": "Z-0380-2018",
      "product_description": "RayStation     Product Usage:  is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.",
      "product_quantity": "117 units",
      "reason_for_recall": "For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.",
      "recall_initiation_date": "20170717",
      "center_classification_date": "20180116",
      "termination_date": "20180829",
      "report_date": "20180124",
      "code_info": "6.0.0.24 ,  6.1.0.26 ,  6.1.1.2  or  6.2.0.7"
    }
  ]
}