      "country": "United States",
      "city": "Maple Grove",
      "address_1": "6250 Sycamore Ln N",
      "reason_for_recall": "Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.",
      "address_2": "",
      "product_quantity": "13 US",
      "code_info": "Serial Numbers:   152806  157499  157683  181193  181364  182887  186654  186767  186797  192099  192182  192351  197342",
      "center_classification_date": "20180126",
      "distribution_pattern": "Worldwide Distribution - US, including the states of :  WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.",
      "state": "MN",
      "product_description": "Metrotom 800 (130kV CT scanner)    Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.",
      "report_date": "20180207",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Carl Zeiss Metrology Inc",
      "recall_number": "Z-0379-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "78885",
      "more_code_info": "",
      "recall_initiation_date": "20171114",
      "postal_code": "55369-6310",
      "voluntary_mandated": "FDA Mandated",
      "status": "Ongoing"
    }
  ]
}