      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78760",
      "recalling_firm": "Roche Diagnostics Corporation",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA (nationwide) Distribution",
      "recall_number": "Z-0378-2018",
      "product_description": "Brand Name:  ¿ cobas c 6000 MODULAR Series System   e 601  ¿ cobas c 8000 MODULAR Series System - e 602  ¿ Roche / Hitachi MODULAR Analytics Combination System    Model/Catalog/Part Number:  MODULAR Analytics E 170  ¿ 03022109001  ¿ 03023109973  ¿ 03739040001  ¿ 03739040692  ¿ 05023572001  ¿ 05023599001  cobas e 601 module  ¿ 04745922001  ¿ 04745922692  ¿ 05036348001  ¿ 05036348692  ¿ 05860652001  cobas e 602 module  ¿ 05990378001      The Roche/Hitachi MODULAR Analytics System (E170) is a fully automated, random-access, software- controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. The cobas 6000 series (e 601) is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module. The cobas 8000 Modular Analyze Series (e602) is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit.",
      "product_quantity": "2517",
      "reason_for_recall": "Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.",
      "recall_initiation_date": "20170629",
      "center_classification_date": "20180116",
      "termination_date": "20200124",
      "report_date": "20180124",
      "code_info": "Serial Number  2072-26  27P6-20  15S2-20  2717-18  2719-02  1544-11  1686-03  1534-12  1698-04  1549-04  2275-12  2263-16  2263-17  2497-16  1849-01  1570-15  26F0-17  2256-18  2564-30  2850-05  2384-12  2384-18  2399-08  2230-19  2230-20  2379-14  2381-02  27Z8-01  2058-22  1909-13  1909-01  27V5-20  15U5-18  2308-01  2087-07  1984-05  1985-07  26D3-30  2497-01  1861-18  1861-19  26D0-27  26D1-21  1480-14  1480-15  1488-01  1488-02  26C3-12  26F0-19  26F0-27  26D9-24  2107-35  2348-12  2348-13  1423-16  2237-18  2240-09  2240-10  16J8-05  16J8-06  16L7-20  16K2-17  16K2-18  16K2-19  16K2-20  16K3-01  16K3-06  16K3-07  16K3-08  16K3-09  16L4-03  16L4-04  26D1-25  26D1-26  26D1-28  26D1-29  2328-03  2437-13  2072-29  2072-30  2064-12  1931-11  2080-35  1936-08  2049-15  27Z0-09  27Z0-10  1911-11  2581-18  2062-14  2074-05  1430-16  1430-17  1691-17  1691-18  1985-13  1989-19  1739-19  2578-25  13B8-12  1169-06  1169-08  1169-09  1211-16  1211-17  1211-18  1224-10  1458-04  2543-10  27J6-15  2048-05  2406-27  26B3-30  1927-19  1914-13  1914-11  27J7-06  2822-07  2822-08  2535-23  2306-04  1159-19  2867-11  2711-02  27Y6-19  2107-20  2195-34  2106-20  2079-34  27M5-13  15D1-02  15C9-10  15F0-17  15F0-18  2137-15  1872-01  2195-40  2102-30  2428-02  2539-08  1377-14  2495-01  1490-16  15C5-16  15C5-17  15C5-20  15C6-01  2250-02  26H6-29  2423-15  2539-06  1377-12  2307-26  2307-27  2227-11  2080-24  1128-05  1128-06  14B1-02  14B2-13  2467-05  27Z8-09  26E4-05  2413-14  2423-19  2086-27  27Z0-14  27Z0-16  2145-16  2321-26  2321-25  26F1-20  26F1-19  1305-40  26G7-23  26G8-09  27L8-12  27X4-14  2719-07  27S8-20  15U1-20  26A4-28  26D1-22  2368-06  1128-19  2427-29  2539-07  1377-13  1357-08  1357-09  1357-10  1344-05  1344-06  14B1-01  1344-02  1344-03  1344-04  1469-03  1469-04  1469-05  1476-20  1482-11  1483-01  1483-17  1483-18  1483-19  1483-20  1484-16  1484-17  1484-18  1490-18  1490-19  1490-20  14A3-01  14A3-02  14A3-03  15B7-01  15B7-02  2430-27  2430-29  15B7-03  2327-25  2347-04  2369-04  2332-04  2332-09  27S4-02  27S4-03  2230-14  2230-18  1211-15  1211-14  15D2-04  15D2-05  15D3-12  15F0-15  15F0-16  16P4-02  16P4-03  2061-21  14A9-16  14A9-17  14A9-18  15B7-04  2256-19  2865-20  2867-01  2070-11  2070-13  2720-17  16X4-17  2102-29  2192-37  2196-02  2535-18  2842-03  2896-04  16P4-04  16P4-07  16P4-08  16P4-09  2311-12  2311-13  1871-09  1125-07  1128-02  2059-28  15H8-11  15H8-12  1996-01  2396-04  2399-06  2068-24  2283-14  2080-02  2577-04  2047-10  2107-05  26E4-10  27J7-04  2585-07  2581-08  2581-13  2327-11  2396-07  2413-01  2413-04  185731  1867-06  2037-07  2423-14  1549-02  27Z8-02  1860-19  2074-25  2087-09  2103-35  1924-15  2192-02  2437-19  2437-20  27Z0-03  2437-17  1913-03  1913-04  2488-05  2402-05  26C3-11  26C9-12  26E4-08  26F3-21  1545-05  1844-14  2822-09  2822-10  1303-31  1303-32  1357-19  1357-20  2402-30  2195-38  2196-01  2080-11  2080-17  2226-17  2439-12  2107-02  1988-03  2086-40  2195-24  2195-06  2195-08  2072-39  2041-12  2047-07  2192-15  2192-08  1736-04  1991-08  1871-06  2037-14  2055-30  2055-35  2873-02  2891-13  1990-05  192506  2894-12  1936-14  1928-19  2848-17  2867-08  2058-21  2058-25  2073-01  2111-04  130833  130834  2850-06  1334-12  1334-13  1334-14  1476-01  1476-02  1476-12  2049-29  2049-30  28E5-14  2347-11  2367-09  27Z8-12  2842-02  192418  1936-09  28A2-10  28AO-16  2208-17  2227-10  15J5-02  15J5-03  1852-15  2256-15  1461-03  1461-04  1476-11  16M7-01  16M7-16  14B3-06  14B3-07  16K3-15  16K3-16  16K3-17  16K3-18  16K3-19  16K3-20  16K4-07  16K4-11  16K4-20  16K5-01  16K5-02  16K5-03  16K5-04  16K5-05  16K5-09  16K5-10  16K7-04  16K7-05  16K7-06  16L4-05  16L4-06  16L4-07  16L4-08  16L7-19  16L5-03  16L7-17  16P1-13  16P1-14  16P1-15  16P2-19  1417-20  2581-25  2587-14  1431-14  2065-12  1545-03  16M7-13  16M8-01  16M8-02  16P1-11  2275-13  2267-19  2275-09  2275-10  2056-14  2047-08  185715  185817  26C3-13  2137-09  15D9-17  15D9-18  2486-24  2489-30  2054-06  2054-10  26E0-18  2467-15  16J8-12  16J8-13  1871-07  2567-22  2367-13  2304-14  2304-15  2202-07  2288-10  1487-16  1488-03  1480-16  1488-04  2427-27  1857-13  2486-06  27J7-03  27Z0-17  27Z0-18  27Z0-19  2494-30  2071-34  2576-22  26E4-06  1124-05  1124-06  1132-03  1132-04  16K5-06  2493-18  2497-12  2848-14  2848-15  2842-01  2848-12  2896-03  2263-13  2263-14  2050-29  2547-29  2550-19  2058-17  2058-20  2054-07  2054-08  2328-05  2439-13  1535-09  2548-03  184826  2497-13  2039-09  2061-35  1905-11  0753-22  1685-16  1570-14  2135-17  2140-11  2493-28  2493-29  2850-09  2850-10  2581-07  2581-22  1867-14  27P6-19  15S2-19  26G7-24  26G7-25  26H3-28  2108-36  192717  2493-23  2494-21  1225-14  2822-20  2824-01  2585-04  1910-02  2540-23  2540-24  1694-19  2104-01  2047-36  0884-01  1991-18  2389-18  2107-29  2108-27  26D6-03  26D5-22  1174-08  2189-02  2189-03  2086-13  2086-21  2446-13  1557-12  2486-23  2867-03  191007  2137-14  16K5-12  16K5-13  16K5-14  16K5-15  2104-02  1725-20  26E4-04  26A9-09  2869-20  16Z2-20  2581-26  2389-13  1142-15  2075-40  0817-20  1927-15  1156-01  1160-04  1156-02  1156-09  1160-06  1331-10  1331-11  1331-12  2072-34  1548-08  2262-18  2405-26  16M8-03  16R5-08  15B7-12  15D2-01  15D2-02  15D2-03  16P2-20  2720-20  2451-02  2451-06  2451-09  2468-18  1985-12  1860-15  2850-03  2850-04  2067-31  2086-28  0829-17  191305  2107-11  1848-01  1848-03  27Z0-15  2704-19  2047-39  0884-04  1421-11  2423-16  2423-17  1482-12  28E7-02  2413-02  2413-03  2486-05  2074-17  2486-04  2486-16  2327-23  2398-05  2438-15  2438-16  2479-09  2479-11  2311-07  2311-11  2425-11  2427-18  2227-07  2233-16  27X4-18  2427-21  2381-03  2382-04  2822-19  2867-07  2850-01  2382-06  2227-02  26C7-30  14H8-04  1017-06  1021-03  1021-05  2108-31  2108-32  2414-14  2367-11  2367-12  2451-19  2406-28  27M5-12  2368-20  2369-01  2398-14  26G4-01  26G4-09  1686-01  2369-05  16S0-01  2080-19  1874-06  26D8-26  26E4-14  26D5-24  26D4-26  26D4-27  2039-07  2585-27  26A4-26  2570-30  13A9-09  2240-13  2240-15  1914-18  2072-31  2072-18  2304-13  2362-19  1122-18  1924-13  2049-14  16R5-15  27L8-11  1868-08  2446-12  2535-20  28E7-16  2548-24  1387-17  1486-20  1490-17  14A3-04  14A3-05  14A3-06  14A3-07  1118-13  2534-05  2534-06  1872-12  2047-37  0884-02  2413-09  2891-15  26H1-28  2858-02  16J8-16  16J8-17  16K2-01  27P2-10  2704-20  2441-10  2441-11  2422-18  2425-12  2219-18  2234-07  1988-11  26B5-23  2379-13  2406-29  2407-04  2894-20  2896-01  1919-05  2102-31  1872-07  1476-03  1479-07  2576-21  185503  27S2-18  2228-15  1005-29  2413-11  2429-30  1263-20  27P2-16  27P2-17  2108-05  2108-21  2425-14  2039-13  2041-16  2413-06  26F1-17  27S4-04  27S4-05  2585-19  2585-21  2493-05  2493-19  1910-04  2422-29  26D6-02  26E0-16  2842-11  2107-31  2848-18  2193-39  27R3-06  27R3-08  27L1-09  2842-10  1914-20  0763-09  2347-15  2348-16  27Z8-08  1691-15  2427-25  2427-26  2207-08  2226-09  26C7-29  14H8-03  2577-02  2577-01  26D4-29  26D4-30  1927-09  2848-16  1459-12  1459-13  1476-05  1476-06  2073-10  26C3-07  190703  2049-21  2369-03  2587-10  2305-22  2070-21  27J7-09  2066-27  2103-40  26H3-26  2848-19  2061-36  0801-27  2581-23  2584-23  2584-30  2395-29  2396-02  1870-16  2073-38  2080-30  2578-24  13B8-11  2071-20  2071-21  2065-04  27W3-18  1344-01  1344-07  1344-08  2288-12  2288-07  27Z8-18  27Z8-19  28E7-13  28E7-14  2140-06  2140-07  2257-02  2079-30  16P1-10  1479-08  1476-07  1476-08  1476-09  1476-10  1988-02  2137-17  14A9-01  14A9-05  2563-30  27R8-19  15S9-19  26B4-26  27J7-02  26H6-26  27Y6-16  27S0-18  27S0-19  26D9-30  2473-07  2486-21  27M5-14  2368-03  1128-16  2319-06  28A3-02  2363-01  26H5-28  28D3-18  1122-20  14M7-18  1616-18  16X4-19  2711-01  2232-19  2708-15  26D6-04  26D8-24  1872-15  1873-05  2054-11  2193-11  2140-03  1855-33  1927-06  1927-20  2074-15  2107-07  2074-14  2209-20  2078-16  2085-33  2137-11  2347-16  1914-17  2358-18  1914-15  0763-06  1118-12  0763-04  15F5-15  15F5-16  2077-38  2078-20  15H2-05  16K4-05  2074-30  2408-22  2580-28  2580-29  1679-13  2497-02  2497-03  26A9-06  1138-10  1139-01  1127-08  1138-08  16M7-05  1988-06  2148-15  28E5-13  2486-27  2493-04  1912-20  0761-08  2374-11  1684-07  1684-08  1570-17  1479-06  1480-17  15F5-12  15F5-13  15H2-08  15H2-09  15D9-15  15D9-16  1691-19  1548-16  2126-08  15H7-15  15H7-16  15H7-17  15H7-18  2068-33  1857-12  2850-15  2106-17  2858-03  2858-04  27P2-07  27P2-08  26D8-20  1418-03  1418-04  1561-12  1925-04  16P4-11  16P4-12  16P4-16  16P4-17  16P4-18  2187-36  2187-37  2187-38  2331-01  2085-09  2086-17  2086-18  2086-15  2086-16  2086-19  2188-08  2188-09  2188-10  2188-36  2188-37  2188-38  2102-06  2102-07  2102-08  2253-16  2253-17  2253-18  2260-11  2260-12  2260-13  26H6-24  2493-27  1858-04  1681-09  1860-21  1861-24  2065-10  1860-20  1860-17  2050-04  2566-26  27P0-20  2850-16  16K6-11  16K6-12  16K6-13  16K6-14  16K6-15  16K6-16  2061-20  2085-12  2085-14  2283-15  2065-06  2081-07  2081-08  2081-09  2081-10  2081-11  2083-04  2083-05  2084-22  2084-23  2083-06  2083-07  2083-08  2188-31  2188-34  2188-35  2192-21  2192-22  2192-26  2192-27  2260-17  2260-18  2260-19  2260-20  2262-11  2262-12  2260-14  2260-15  2260-16  2071-25  2075-20  2263-15  2283-18  2451-17  2451-18  2347-19  2348-11  28E5-12  1733-15  2052-01  2052-02  16K6-17  16K6-18  16K6-19  2086-25  2086-26  2189-07  2082-39  2084-05  2084-12  2084-13  2084-14  2084-15  2084-06  2084-07  2192-31  2101-39  2257-08  2257-09  2257-10  2257-11  27S2-05  27S2-06  2050-06  2050-07  15D6-18  15D9-12  15D9-13  15D9-14  2071-22  2086-23  2086-24  2252-05  2255-20  2263-12  2084-08  2084-09  2084-10  2084-11  2085-04  2085-10  2188-04  2187-33  2188-39  2188-40  2102-03  2195-23  2146-17  2146-18  2146-19  2253-19  2253-20  2254-01  2257-04  2577-03  16K5-17  1616-17  27Z0-07  1905-05  1307-41  2050-01  1544-13  1536-04  2085-11  2084-19  2084-20  2084-21  2413-08  2891-14  2446-14  2446-16  1537-15  1989-20  2050-08  15H7-11  15H7-12  15H7-13  2082-40  2083-01  2084-16  2084-17  2084-18  2188-32  2189-37  2189-38  2189-39  2251-16  2251-17  2251-18  2452-01  2469-06  2232-18  2234-02  26C3-08  26C3-09  2488-02  2232-20  2039-10  1357-15  1357-16  1357-17  1357-18  26G8-08  27R8-20  15S9-20  2493-13  2493-15  2469-05  2469-07  2234-01  2348-17  2145-17  2850-12  2850-13  2145-11  2145-12  15D3-13  15D5-04  15D5-06  15D6-17  1308-35  1308-36  1419-12  1422-17  16J8-14  16J8-15  2079-18  2081-12  2192-18  2147-05  2147-06  2085-05  2085-07  2085-08  2190-14  2188-01  2119-14  28D7-19  1616-19  1921-01  1921-02  1537-17  1905-01  1905-02  2050-03  2050-05  15E3-04  15E3-05  15E3-06  15E3-01  15E3-02  15E3-03  2085-16  2085-17  2189-01  2189-05  2189-06  2256-01  2278-09  2081-06  2188-05  2188-06  2188-07  2190-17  2190-18  2190-21  2190-03  2190-04  2190-05  2189-40  2190-01  2190-02  2189-36  2192-24  2086-12  2106-28  2106-29  2110-09  2110-10  2188-02  2119-13  2056-21  2147-04  2251-15  2253-12  2253-13  2253-14  2257-05  2257-06  2257-07  2059-05  2140-08  2140-10  2473-02  2473-03  2076-22  1544-10  1733-18  1725-19  1857-27  1556-12  1681-08  1694-20  16M6-20  16M7-02  16M7-03  16M8-04  2076-19  2076-20  2076-21  2083-03  2085-18  2083-02  2085-13  2275-14  2275-15  2275-16  2278-07  2190-13  2190-15  2190-16  2192-28  2192-29  2290-05  2290-06  1711-16  2192-12  2189-04  2086-14  2086-20  2190-10  2190-19  2190-20  26E4-12  26H6-23  27L7-20  2069-01  2087-03  26D8-21  26D8-25  2873-01  2891-18  1921-03  1557-10  1736-07  1855-12  1905-03  1925-03  16J8-07  16J8-08  16J8-09  16J8-10  16J8-11  2080-03  2085-15  2252-04  2252-03  2278-08  2079-19  2080-01  2081-04  2190-06  2190-07  2190-08  2190-09  2190-11  2190-12  2192-25  2192-32  2252-02  2253-15  2251-19  2251-20  2283-16  2296-20  27S8-10  27S4-06  2486-14  2486-15  27R3-20  27R5-01  2869-18  16Z2-18  2107-04  2485-29  2487-28  2587-12  2145-08  2290-12  2486-02  2062-02  2467-03  2066-22  2066-23  2102-36  2708-19  2719-08  27R3-05  1556-16  1861-23  2581-21  2066-35  2066-36  26G4-18  26G4-11  26G7-22  26C9-13  26C9-14  2070-15  2472-10  2481-09  2195-36  1160-01  1160-05  1183-04  1217-09  1225-11  26E4-11  2212-19  2212-20  1864-17  1864-18  2427-13  2431-26  26D6-06  14A8-02  16L5-02  16L8-01  16L8-02  16L8-03  16P4-10  27M5-15  2399-09  2399-13  2399-04  2087-08  2585-18  1459-10  1459-11  2104-04  2104-07  2472-11  2472-12  2423-18  27X6-02  27X6-03  27X6-04  27J7-10  27S2-17  27S2-20  2493-26  2212-12  2212-13  2232-13  2232-14  2232-15  26E0-19  2145-13  26B5-24  26C3-05  26C3-06  2107-30  2137-12  2137-13  2581-16  2441-04  1684-05  1572-19  2192-10  2103-29  2104-06  2356-01  27P0-08  27S0-16  1571-13  1867-19  1935-18  1908-03  0756-30  1139-02  1139-03  1139-04  1139-05  2427-19  2427-20  2497-14  2305-23  2307-04  27S2-19  2543-04  27Z0-04  2059-23  2382-05  26G6-30  2358-14  2570-29  1118-08  13A9-08  2348-20  26E4-07  1684-06  2057-28  2487-30  1873-03  2535-22  2535-21  2049-16  27S0-20  186225  2395-28  2865-17  2533-21  1910-08  27S4-01  27P2-01  2485-30  26D8-23  26H6-25  2367-10  2439-14  2486-18  2389-15  1142-17  1860-22  2389-12  1142-14  2865-19  1310-20  1871-12  2069-24  2250-08  2256-14  27S2-01  2227-14  2547-30  2548-01  1925-07  2413-15  2565-21  2565-22  2577-05  2585-22  1556-17  1843-02  26F0-29  27L8-01  27L8-08  2409-26  26A4-27  26B6-30  14G5-06  2565-23  2565-28  2073-27  2079-39  2427-23  2427-24  2544-21  1383-03  2566-25  2566-30  2486-07  2402-28  2226-16  2226-18  26D6-11  26D8-28  2585-25  27M5-11  26A4-23  26A4-29  26A5-01  2587-13  26D0-25  26D0-26  26D5-30  26D9-25  26E0-14  26G4-13  2585-20  2432-29  2432-30  2432-24  1211-19  1211-04  1211-05  1335-08  1335-09  1335-10  1335-11  1910-06  2438-14  27L1-05  2077-40  2078-21  2535-12  2535-13  2535-14  2550-12  2550-14  2565-25  2565-26  27J6-19  2550-15  2564-23  2565-24  2108-30  14B1-03  14B1-04  2423-03  26B4-22  2072-40  2107-38  2229-12  2233-20  2450-17  2451-01  2446-17  2446-19  2406-25  27Z0-12  27P6-16  27P8-01  16R0-06  16R5-06  2581-14  2581-15  2706-01  2493-25  2867-09  2871-19  27R1-16  2355-01  2485-28  1931-09  2039-11  2224-17  2473-04  2473-06  2486-17  2714-09  2714-10  2458-23  1387-16  26F3-20  26G4-12  2581-17  2437-18  27J6-17  2486-13  27Z6-18  27Z6-19  2472-08  2472-09  27S2-15  2720-16  16X4-16  26G4-10  16K3-11  16K3-12  16K3-13  26H3-25  2427-16  2427-17  26H5-29  14M7-19  2396-18  2321-23  2321-24  2362-20  1122-19  1146-01  1146-02  1146-03  16K3-10  2071-40  2219-17  2229-14  2535-06  2535-07  27W3-19  27W3-20  2398-15  2398-16  2407-06  26D4-28  26D8-22  2256-17  26D0-21  26D1-30  2061-29  1139-06  1139-07  15H2-11  15H2-12  15H2-14  2585-06  28E7-12  28F7-02  1132-05  1132-10  1137-02  2423-21  2425-13  26H0-30  14M0-20  2350-04  2408-23  2409-23  2432-25  1924-12  1935-17  2073-37  2042-03  2497-11  2140-12  2704-16  27L1-15  2220-06  2444-07  26F0-18  2080-25  2106-10  28E9-12  28F5-13  16R5-07  14B1-07  15B7-07  15B7-13  2842-14  2842-15  1853-03  1860-16  1733-16  1736-03  2070-01  2070-02  1601-03  1705-20  1208-05  1211-07  1211-08  2330-20  2409-24  2409-25  2107-18  1139-08  1139-10  1140-07  1140-10  1140-08  1156-06  1158-09  1156-07  14B3-08  14B3-09  14B3-10  2369-02  2564-21  2564-22  2227-09  1572-20  1673-01  27Y6-11  27Z0-06  27Z0-11  2413-12  2413-13  1928-01  1848-04  1925-08  1909-09  1909-10  1926-03  2567-23  2423-05  1936-10  1171-10  1211-13  1988-12  1983-06  2494-22  2054-09  1563-04  26B4-21  16K7-12  16M7-06  2856-19  2869-17  2856-20  16Y8-19  16Z2-17  16Y8-20  16L3-20  16K7-11  1866-13  2427-30  2428-01  26A5-08  26A9-03  2304-30  2305-08  2137-06  2564-29  1984-01  2384-19  26A4-24  26A4-25  2048-06  2048-29  2050-28  2072-19  26D6-07  26D6-08  26D6-10  15F0-14  15H2-06  15H2-07  15H5-01  2423-22  2867-05  2867-06  2894-07  2867-12  2894-10  2894-11  16K5-16  16K5-20  16M7-07  16M7-08  2867-13  2585-05  2585-26  2145-14  2145-15  2080-26  27X4-19  27X6-01  2398-17  2396-01  26A9-07  15H4-08  15H4-18  15H5-02  15H5-03  16K5-17  16K5-19  2069-08  2070-05  2313-17  2871-20  16R5-09  16R5-10  16R5-11  27L8-13  1992-19  1999-16  1999-17  2550-10  2550-11  1865-05  1908-10  0757-07  2867-10  16P1-06  16P1-07  16P1-08  16P1-09  2306-02  2306-03  1159-17  1159-18  27P2-05  1927-08  2219-20  16J9-20  1611-20  2312-26  2312-28  2444-08  1987-12  28F5-17  2493-08  2497-17  2422-17  26G4-16  2137-16  2896-02  27P2-04  2079-27  2079-36  2565-27  2567-21  2074-04  2074-01  2538-29  15B7-05  15B7-09  15B7-10  26B3-29  1991-07  1860-23  1871-11  2059-11  27L1-14  27Z8-13  2550-04  2896-05  2060-25  2066-39  26D5-28  16M7-19  16M7-20  1173-01  1985-06  2842-13  27Z8-04  27Z8-05  1919-07  1919-08  2560-03  13A1-02  2256-16  1919-03  1911-15  2050-31  2050-32  1909-08  2074-13  26G4-15  27Y6-15  27Y6-18  2842-17  2858-01  2188-03  26G4-17  2497-22  2538-28  1156-05  1158-10  1461-01  1461-02  1870-06  1870-14  2429-29  1263-19  2071-24  1469-01  1480-18  1925-12  1926-02  1548-17  1845-15  198703  1987-04  1159-20  15D9-20  15D9-19  2145-09  2145-10  2148-16  27Y6-10  27Z0-05  2218-14  27J7-08  2869-19  16Z2-19  2894-14  2137-18  2149-14  2585-24  2384-20  2385-01  27X6-05  2347-09  2347-10  2358-17  1118-11  26C3-14  26C3-20  2708-17  2708-18  2234-03  1156-08  1156-10  1344-09  27Y6-17  27S0-17  27Y6-13  27X4-15  26H4-29  27S2-11  27S2-12  2066-38  1867-20  1868-01  2047-09  2086-35  2369-06  2720-18  16X4-18  1987-14  1874-05  1985-08  1985-10  2494-28  2497-21  2064-06  2067-29  2472-19  2321-21  2195-03  1714-18  26B4-24  26B4-25  2495-02  2543-03  2543-08  2543-09  26E0-22  27Z0-13  2048-32  27S2-02  27S2-03  2704-18  2719-03  1208-08  1208-09  1211-02  1211-03  2367-14  2367-15  2059-10  27S2-04  27S2-08  27R5-02  27R3-18  2051-32  2896-06  2894-13  2058-32  27Y6-12  27S2-13  2102-33  2219-19  2225-01  2234-06  2224-15  2224-16  15C9-09  2229-10  2233-19  2233-13  2234-04  2496-21  1871-03  2382-07  2382-12  2403-21  2493-24  26H3-27  2848-11  26D8-19  26D8-27  2479-08  1601-02  1858-05  2317-18  2332-11  2332-12  2396-13  2398-06  2312-27  2332-20  26D0-24  1853-02  1855-10  1855-11  1858-02  1858-03  1849-02  1853-01  1691-14  1860-12  1860-13  1860-14  1860-11  2104-21  2104-03  2368-04  1128-17  2070-22  2581-19  2581-20  27J6-20  27X4-16  27X4-17  2850-07  2850-08  1146-05  1146-06  1147-04  1485-11  2067-30  2070-40  2414-10  2414-11  2414-12  2427-22  2430-28  2431-24  2431-25  26D9-29  26E4-09  2850-11  28B6-19  1611-19  2192-16  27Z8-10  27Z8-11  26G4-19  2858-05  2860-20  2368-13  2369-13  2304-12  1131-08  1132-01  1132-02  15J6-10  2144-13  2140-04  2894-09  2894-16  2894-17  1983-09  27Z8-15  2822-11  28D7-20  28F5-18  2432-27  1862-27  2719-09  2304-29  2305-09  2566-28  2566-29  2073-09  2059-40  2422-23  2422-24  2149-13  2430-30  2431-21  2431-22  2108-25  2108-26  1988-14  2494-27  2209-09  2209-10  26B6-29  14G5-05  2072-28  27J7-05  26G7-21  2073-40  27Z8-17  2848-13  26F0-30  2534-04  2534-03  2535-09  2535-10  2535-11  2405-28  2405-29  26B3-28  2544-22  2550-18  2488-04  27W5-01  1338-12  1338-13  1338-14  2202-11  2292-17  143115  2209-19  2327-26  1988-10  2056-09  2056-12  1559-13  1539-13  1539-14  1571-12  1480-19  1490-09  1490-13  1490-14  1491-01  15B7-08  28F5-14  2468-17  28E7-15  2432-22  2432-23  190520  2149-12  2104-08  2104-20  1867-16  1156-04  2444-14  1868-02  2423-04  2535-05  2192-13  2196-03  2497-15  2422-16  2137-03  2137-04  1698-05  1711-17  1991-13  1991-14  2233-17  2233-18  2543-11  2543-12  1544-06  26F1-18  26H6-28  26G4-14  2071-23  27Y6-14  2422-28  2423-20  2056-10  2056-13  2581-24  2585-23  2086-36  2108-37  2047-38  0884-03  16L7-18  16P4-01  2085-30  2087-11  26F0-28  26F1-16  27X8-20  15V4-20  2321-22  16P1-12  16P3-02  16P3-03  1865-19  1999-20  1861-20  15H7-19  15H7-20  15H8-01  15H8-09  15H9-13  15H9-14  15H9-15  15H9-16  15J0-03  15J0-04  15J0-06  15J0-08  16K4-01  16K4-02  16K4-03  16K4-04  16K5-07  16K5-08  2267-20  2262-19  2262-20  2221-03  199902  2414-09  2535-08  2385-02  2385-03  2327-12  2867-02  2427-14  2427-15  2234-08  2234-09  2236-10  1008-10  16K6-20  16K7-01  16K7-02  16K7-03  2486-12  2389-16  1142-18  1461-05  1468-18  1468-19  1468-20  2330-16  2330-18  1171-07  1171-08  1171-09  1173-10  1160-08  1171-06  1174-06  1174-07  2275-11  2278-10  2278-11  2330-19  2140-02  2262-17  2267-17  2267-18  2283-17  2330-17  2585-01  2585-02  2585-03  2585-16  2585-17  2399-05  2399-07  2233-15  2891-19  2079-29  15G9-03  15G9-04  2494-29  26G3-30  27J7-12  2305-10  1142-09  1145-10  26E0-20  26E0-21  2293-05  1140-01  2052-05  26D5-23  2196-07  1858-35  26D0-22  26D0-23  2192-09  2322-08  2107-22  2107-23  2190-38  2195-10  2402-29  2195-01  2103-36  2321-29  2066-28  1139-09  1173-08  2070-07  2310-08  1334-15  1335-13  1459-14  14B2-14  2488-03  1428-07  1182-08  1211-06  1331-13  1338-15  1338-16  27S2-14  1147-05  1149-01  26C7-28  14H8-02  1926-04  2432-26  2432-28  2566-27  26E4-13  15J6-02  15J6-08  2427-28  2037-13  2059-12  2058-18  2429-28  1263-18  2148-14  2149-08  2149-20  16P6-19  16R7-03  2256-02  2076-18  2262-16  2414-13  2891-16  2891-17  2148-09  28E7-18  2086-31  0829-20  2894-18  2894-19  1601-01  1698-03  16P6-14  2058-23  1429-13  2250-01  2307-05  1919-02  1919-04  2220-02  2220-04  1469-02  2220-07  26A5-02  2850-14  2431-23  2471-20  27K0-06  1211-20  15B7-06  27R3-09  27R3-13  1907-02  2321-20  2438-04  2409-22  2408-20  15C8-11  14B1-05  14B1-08  2332-10  2438-13  26E0-15  2037-11  1991-17  1991-16  26H6-27  2061-40  0802-01  199010  27Z0-08  15R8-20  27J6-14  27J6-13  27J7-01  27J6-18  27Z8-07  27P2-09  2479-10  2493-17  26D5-21  2850-02  27X6-06  27Y6-09  187018  1925-13  27Z8-06  2488-01  26F1-21  2708-16  27Z8-14  2042-02  2321-27  26A9-10  27L1-12  27L1-13  27L1-03  27L1-04  27K9-16  27L1-01  27L1-02  1983-08  1868-14  2288-08  2328-02  2473-01  2077-36  1928-02  1924-11  2192-11  1149-07  1149-08  1156-03  1167-02  16R0-07  16R0-08  1863-05  27J7-11  27J9-12  27L1-06  1935-16  26G4-08  15C9-08  27L1-07  27L1-08  15C8-12  15C8-13  15C8-14  15C8-15  27L1-10  27L1-11  2301-05  2308-06  2229-03  2229-07  2067-33  27S2-16  27Z8-16  1698-06  1733-17  26D5-27  26E0-17  1862-16  27J6-16  1538-10  27Z2-09  2494-23  2494-24  2494-25  2494-26  28F5-15  28F5-16  16R5-04  16R5-05  2061-34  2064-05  2137-05  2227-12  2227-13  2413-10  2422-15  2226-19  2233-14  1016-05  1017-01  2102-04  2102-05  27R3-14  27R3-16  2578-23  13B8-10  2708-20  27Z6-20  2867-04  2486-22  16L8-08  16M4-14  16M4-15  16M4-16  16M7-04  2550-16  2563-29  1905-06  2848-10  2842-04  2061-37  2061-38  0801-28  0801-29  2413-07  2422-19"
    }
  ]
}