  "results": [
    {
      "status": "Ongoing",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91103",
      "recalling_firm": "Bunnell, Inc.",
      "address_1": "436 W Lawndale Dr",
      "address_2": "N/A",
      "postal_code": "84115-2917",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution: PA, NY, AR, AZ, TN, SC, MA, FL, CA, TX, MD, VA, NV, IL, IN, IA, HI, KY, LA, WI, GA, OH, MI, MO, MN, NJ, MS, NC, ND, NM, UT, OK, WA, RI;  OUS: Canada",
      "recall_number": "Z-0371-2023",
      "product_description": "2.5 mm LifePort Endotracheal Tube Adapters 10-pack, Catalog: 9025; and 2.5 mm LifePort Endotracheal Tube Adaptors inside Patient Circuit Kits, Catalog: 938",
      "product_quantity": "1606 Assemblies",
      "reason_for_recall": "Some adaptors, for use with high frequency ventilators, were packaged improperly, and may deform at the tip, which may impair the ability to pass a suction catheter or may affect normal ventilator performance, which may lead to hypercarbia, hypoxia, and gas trapping.",
      "recall_initiation_date": "20221102",
      "center_classification_date": "20221202",
      "report_date": "20221214",
      "code_info": "Catalog/UDI/Lot: 9025/10616120000034/22042320 ; and 2.5 mm LifePort Endotracheal Tube Adaptors, Lot: 22042320, inside 938/10616120000089, 00616120000082/22052381, 22052407, 22062431, 22062458, 22062478"
    }
  ]
}