      "country": "United States",
      "city": "Cincinnati",
      "address_1": "675 Wilmer Ave",
      "reason_for_recall": "The floor locking cylinders contained in the base of the Schaerer 7300 model surgical tables were manufactured with the incorrect \"O\" ring which may cause the cylinder to leak.",
      "address_2": "",
      "product_quantity": "4 surgical tables with defective floor locking cylinders were distributed. Each table contains  3 cylinders for a total of 12 cylinders.",
      "code_info": "Lot Codes for the four (4) model 7300 Surgical Tables   TFX-1653, TFX1654, TFX1655, TFX1656",
      "center_classification_date": "20161108",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA,  and VA    Replcement cylinders wre distributed",
      "state": "OH",
      "product_description": "The surgical table is a stand alone unit. Three (3) floor locking cylinders are attached to the base of the surgical table. The individual cylinders are replaceable.        Product Usage:  The cylinders are used to lock the surgical table in place.",
      "report_date": "20161116",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Schaerer Medical USA Inc",
      "recall_number": "Z-0371-2017",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "75288",
      "termination_date": "20200717",
      "more_code_info": "",
      "recall_initiation_date": "20160927",
      "postal_code": "45226-1802",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}